9.6 GENERAL CONFIGURATION PAGE
The general page configuration, able the user to specify general STAC2 parameters as:
-
Flow cell optical size (2, 5 or 50mm)
-
Measurement cycle frequency in minutes
-
Automatic cycle measurement start in case of power failure.
-
Date and Time
-
Fixed Wire Ethernet connection for MODBUS & remote control
-
WIFI SSID/Password for remote control.
9.6.1 GENERAL PARAMETERS
Flow cell size selection
Measurement cycle frequency
Automatic
cycle
measurement start in
case of power failure
Flow cell size factor :
In case of 2mm or 5mm
Flow cell size selection,
real optical path length
can
varies
around
nominal path length;
because
of
gaskets
compression factor and
mechanical machining
effects. As a result, a
linked factor to the Flow
cell
itself
must
be
entered
(check
this
value on the flow cell’s
bottom side and enter it
there).