STAC2
- 20 -
5.4 SAMPLE INLET
5.4.1 PERISTALTIC PUMP RATE OF FLOW
Around 13 liters per hour
Around 220 ml per 60 seconds.
Theoretic volume = 0.019 liter per tube length meter (tube internal diameter: 5 mm).
5.4.2 SAMPLE CHARACTERISTICS
Generally the sample does not need to be prepared before the measuring.
NOTE 1
Nevertheless, in certain conditions, it must be cooled or warmed to avoid the condensation on
the measure flow through cuvette.
Minimum temperature =10°C Maximum temperature = 40°C
NOTE 2
With the installed tubes, it is advised against passing samples containing concentrated solvents,
oils, concentrated acids and concentrated NaOH.
5.4.3 TAKING SAMPLE CONSTRAINTS
The peristaltic pump can bring a sample from a distance of
5 meters maximum
and a depth of
2 meters
maximum
(see drawing under).
.
Fig. 6 STAC positioning.
5 meter max
2 meter max
FLOOR
1 meter
0.3 meter
0.4 meter
0
.5
m
e
te
r