STAC2
- 54 -
WARNING:
It’s not allowed to set more than one time the same parameter name
even if coming from different model.
Correction factor able customer to apply
a multiplication factor to the calculated
result.
A linear compensation can correct
imperfection in flow cell exact internal
optical size versus theoretical one for
example.
Sensibility level is
the low limit below
which the result is
considered as too
low to be correctly
evaluated /
measured.
An event is
recorded in the
STAC2 events
database each
time result is below
this limit
Alert is the high
limit over which the
result is
considered as too
high for the
application.
An Alert event is
recorded in the
STAC2 events
database each
time result is over
this limit
Restitution Alert, is
the maximum value,
the user can defined,
which represent a
confident level below
the measured value
can be considered as
good.
See Restitution value
explanation for more
details.
An Alert event is
recorded in the
STAC2 events
database each time
the restitution level
above this user
defined limit.