XR-3 TS2 Module Operation Manual
XR-3
Page 61 of 66
6) Attempting to turn on a unit when the battery contains insufficient charge may result in
beeps and/or a flash on the LCD but will not harm the unit.
AC Power
When 110VAC or 220VAC power is applied via the external AC to DC switching power supply, it
supplies 14VDC power to charge the battery, run the meter, and power the LNB. In this case, the
power available to the LNB is limited to 750 mAmps to avoid overdriving the power circuit and the
battery charging circuit.
DC Power
When external 12VDC power is provided from a vehicle, it supplies enough power only to charge
the battery.
16. Troubleshooting / FAQs
a) I receive an error message stating: “Error: LNB under voltage, over current, or short”
i) Replace top (Signal In port) F-connector barrel. Use a 7/16" wrench to loosen
existing barrel. Install new F-connector barrel and only tighten to 20 inch pounds of
torque with wrench after finger-tightening. Warning: Do NOT over-tighten!
ii) Ensure that you have the latest available version of Field Guide and Software loaded.
If you don't, you must update meter's software using the FlashUpdate program.
iii) If issue still remains, contact Applied Instruments' Tech Support department and
request a RMA# to send the instrument to the factory for inspection and repair.
b) While I have the LNB connected to the ODU / LNB port of the meter with a coax cable
and the LNB power is turned on, I don't even see any signal level shown on left bar
graph. It just says "UNDER".
i) Replace top (Signal In port) F-connector barrel. Use a 7/16" wrench to loosen
existing barrel. Install new F-connector barrel and only tighten to 20 inch pounds of
torque with wrench after finger-tightening. Warning: Do NOT over-tighten!
c) While attempting to perform a Limit Scan, I receive an Error message saying "ERROR:
Key transponder not found or invalid".
i) Turn the meter off and on. Then go to MENU/OPTIONS and press the PRESET soft-
key in the lower left corner of the LCD screen. Then exit back to the MAIN MENU
and try to run your Limit Scan again.
d) I believe my dish is properly aligned, but I'm not getting a LOCK on the meter for the
desired satellite.
i) Ensure that you have the latest available version of Field Guide and Software loaded.
If you don't, you must update meter's software using the FlashUpdate program.
ii) Ensure that you are using the proper LNB type and that you have specified the
correct LNB type in the System Setup software menu. If in doubt, refer to the label or
tag on the body of the LNB and contact Applied Instruments' Tech Support
department to consult with them.
iii) If the incoming signals are linearly polarized, you must ensure that you have properly
adjusted the polarization offset of the LNBF or feed horn assembly. Even if the
antenna reflector is set to the proper azimuth and elevation, it is possible to not
receive adequate signal if your polarization offset is very far off.
iv) If you're working with circularly polarized signals on a multi-LNB dish assembly, you
must ensure that the skew/tilt of the dish is set correctly. You can use the ZIP postal
code lookup feature to obtain approximate tilt/skew settings.
TESCOM USA
TESCOMUSA.COM
Содержание XR-3
Страница 11: ...XR 3 TS2 Module Operation Manual XR 3 Page 11 of 66 TESCOM USA TESCOMUSA COM ...
Страница 39: ...XR 3 TS2 Module Operation Manual XR 3 Page 39 of 66 TESCOM USA TESCOMUSA COM ...
Страница 50: ...XR 3 TS2 Module Operation Manual XR 3 Page 50 of 66 TESCOM USA TESCOMUSA COM ...
Страница 58: ...XR 3 TS2 Module Operation Manual XR 3 Page 58 of 66 TESCOM USA TESCOMUSA COM ...