Adjusting the Baseline
Data Explorer
™
Software User’s Guide
5-55
5
Troubleshooting
If the baseline rises preceding and following a peak after the
correction (a “hump” under the peak), adjust the following
parameters in the order listed:
•
Decrease Flexibility
•
Decrease Degree
•
Decrease Peak Width
NOTE: Lower Peak Width values increase the time needed
for processing.
Returning to the
original spectrum
To return to the original trace, see “Returning to the original
spectrum” on page 5-3.