Overview
Data Explorer
™
Software User’s Guide
3-5
3
Overlapping peak
detection ranges
To accommodate spectral peaks that occur on the boundary of
two peak detection ranges, the software creates detection
ranges that overlap (Figure 3-2).
Figure 3-2 Overlapping Detection Ranges
If a peak occurs in an overlapping range (for example, Peak B
in Figure 3-2), the software detects the peak using the settings
from the detection range within which the entire peak occurs
(Range 2 in Figure 3-2).
If a peak occurs completely within two ranges (for example,
Peak D in Figure 3-2), the software uses the settings from the
higher m/z detection region (Range 4 in Figure 3-2).
PSD peak
detection for
Voyager data
For PSD data, the mass-time correlation is different from that
of standard TOF spectra. Therefore, resolution-dependent
settings are not supported for PSD data.
Range 1
Range 2
Range 3
Range 4
Peak A
Peak B
Peak C
Peak D
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...