![Applent Instruments AT5220 Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/applent-instruments/at5220/at5220_user-manual_2954884033.webp)
33
SCPI Command Reference
9.4
DISPlay SUBSYSTEM
DISPlay subsystem is used for switching different display page or display a string of text on
page reminder bar.
Figure 9-2 DISPlay SUBSYSTEM tree
DISPlay
:PAGE
{MEASurement, SETUp,COMParator,SYSTem,SYSTEMINFO(SINF)}
:LINE
<string>
9.4.1
DISPlay:PAGE
DISP:PAGE is used for switching to an appointed page.
Command Syntax
DISPlay:PAGE <page name>
Parameter
<
page name
> includes
:
MEASurement
<MEAS DISPLAY> page
SETUp
<SETUP> page
COMParator
<COMP> page
SYSTem
<SYSTEM CONFIG> page
SYSTEMINFO
(
SINF
)
<SYSTEM INFO> page
Example
SEND
> disp
:
page
setup
//switch to Setup page
Query Syntax
DISP:PAGE?
Query
Response
<page name> abbreviation
meas
setu
syst
sinf
comp
Example
SEND
> disp:page?
RET
> meas
9.4.2
DISP:LINE
DISP:LINE Is used for display a string of text at the bottom of reminder bar page, text can
display maximum 30 characters.
Command Syntax
DISPlay:LINE <string>
Parameter
<string>
maximum 30 characters.
Example
SEND
> DISP:LINE “This is a Comment.
”
9.5
FUNCtion SUBSYSTEM
Figure 9-3 FUNCtion SUBSYSTEM tree
FUNCtion
:RANGe
{Range NO.
,
max
,
min}
:MODE
{AUTO,HOLD,NOMinal}
:RATE
{SLOW,MED,FAST,ULTRA}
:SCAN
{<
Channel NO.
>,on,off}
Parameter set by FUNCtion SUBSYSTEM will not be saved in the instrument system, users
need to reset when restart.
9.5.1
FUNCtion:RANGe
FUNC:RANG is used for set up range mode and range NO.
Command Syntax
FUNCtion:RANGe {<Range No.>,min,max}
Parameter
Of which
,
<Range NO.> include
1~5
Содержание AT5220
Страница 24: ...24 AT5220 User s Guide ...