![Applent Instruments AT5220 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/applent-instruments/at5220/at5220_user-manual_2954884028.webp)
28
AT5220 User
’
s Guide
8.
Remote Control
This chapter provides the following information to remotely control the AT5220 via the RS-232C
or USB interface.
About RS-232C
About RS-232C connection
Select Baud Rate.
About SCPI
AT5220 can use the RS-232 interface to communicate with the computer to complete all the
instrument functions.
8.1
RS-232C
RS-232 is currently widely used serial communications standard, is also called asynchronous
serial communications standard, it is applied to realize communication of PC and PC
、
PC and
peripheral. RS is the English abbreviation for
“
Recommended Standard
”
(recommended
standard), 232 is standard number, this standard is officially announced by EIA in 1969.
Most configuration of serial port is not based on RS-232 standard: each port use 25-core or 9-
core connector (now all PC use 9-core connector). The most common RS-232 signal is as
below:
Table 8-1 RS-232 Standard minimum subset
Signal
Mark
9-core connector Pin No
Transmit Data
TXD
2
Receive Data
RXD
3
Ground
GND
5
8.1.1
RS232C Connection
RS-232 serial interface can be connected to serial interface of controller (such as PC or PLC )
by DB-9 cable.
Tip: instrument cannot use null modem cable.
Users can make it or buy 9-core cable from Applent Instruments.
If users make 3-core cable, should pay attention to:
If using PC’s built-in DB9 port, probably users need to short circuit 4-6
、
7-8 on PC
port’s DB-9 connector (pin).
Figure 8-1 Connector at rear panel
Содержание AT5220