AT5220 User
’
s Guide
Procedure to perform short-circuit clear zero
Step 1
Press
【
Setup
】
to enter Setup page
Step 2
Short clamp test clip
Step 3
Use function key to select
【
Clear
】
Step 4
Press
【
OK
】,
instrument starts to perform clear zero
Step 5
Return to Setup page after the process is completed
5.3
Comparator Setup
Press
【
Setup
】
key
,
then press
【
Comp Setup
】
to enter < Comp Setup > page
All channels can be set upper limit and lower limit by comparator.
To take full advantage of comparator, instrument adopts built-in Handler interface.
In <Comp> page
,
users can set
:
【
Comp
】
switch
Compare
【
Mode
】
setup
【
Output
】
signal setup
Input data of
【
upper limit
】、【
lower limit
】
for each bin
Figure 5-3 <Comp Setup> Page
Independent Comp Set
Identical Comp Set: