16
AT5220 User
’
s Guide
INT
Continuously repeats the measurement cycle.
MAN
Performs one cycle of measurement each time you press the [Trig] key.
EXT
Performs one cycle of measurement each time a rising pulse is input to
the handler external trigger input pin on the rear panel.
BUS
Performs one cycle of measurement each time it receives a trigger
command sent via RS-232.
Procedure to set up trigger:
Step 1
Press [Meas] key to enter <MEAS DISPLAY>Page
Step 2
Use cursor keys to select [TRIGGER] field
Step 3
Use function keys to select trigger mode
Function keys
Function
INT
Automatic Internal trigger
MAN
Manual Trigger by pressing [Trig] key
EXT
External Trigger Mode
BUS
BUS Trigger by RS232 SCPI command
5.1.2
【
Range
】
SCPI command
:
FUNCtion:RANGe {<Range NO.>,min,max}
5 ranges are available for AT5220, changes for each range is as below:
AT5220 only use manual range to make measurement:
Table 5-1 Range
Range
NO.
Range
Range Ability
UP
DOWN
1
100m
Ω
29.000m
Ω
~300.00m
Ω
300m
Ω
3
Ω
30
Ω
300
Ω
290m
Ω
2.9
Ω
29
Ω
290
Ω
2
1
Ω
290.00m
Ω
~3.0000
Ω
3
10
Ω
2.9000
Ω
~30.000
Ω
4
100
Ω
29.000
Ω
~300.00
Ω
5
1k
Ω
290.00
Ω
~3.0000k
Ω
Procedure to set up Range
:
Step 1
Press
【
Setup
】
key to enter Setup page
;
Step 2
Use cursor keys to select [Range] field
;
Step 3
Use function keys to select range
Function keys
Function
In
Increase range
,
range is changed to LOCK
Decrease-
Decrease range
,
range is changed to LOCK