background image

 

 

 

 

23 

Knobs 

Correct fixing 

 

 

 

Worn out parts, such as the hinged 
points 

Damages 

 

 

 

II. Electrical measurement in accordance with VDE 0751-1:2001-10 (power from the device, direct ) 

Proceed as follows: 

ok 

not 

ok 

Problem 

description 

1. Insert the power cord of the bed into the testing outlet of the measuring 

device. 

 

 

 

2.  2. Connect the probe of the measuring device to a free conductor 

element of the frame (such as a screw or similar).  

 

 

 

3.  Activate the motors with the hand control for the duration of the 

measurements 

 

 

 

4. Start the measurement procedure with the appropriate device.  

 

 

 

 

Limit 

Value 

Value of 

the first 

measure

ment 

Current 

actual 

value 

 

 

 

Result: bed SK II (type B) 

0,1 ma 

ma 

ma 

 

 

 

 

 
Protocol system for testing electrical medical devices 
Sheet 2 to 2 

 

III. Functional control 

ok 

not 

ok 

Problem 

description 

Functional control of electrical components 

End stroke of the motors 

Automatic  disconnection  

 

 

 

Controller and motors 

Unusual noises 

 

 

 

Controller and motors 

Test according to use instructions 

 

 

 

Strain relief cord 

Connecting cable properly fixed  

 

 

 

Functional control of the mechanical components 

Hinged and pivot points 

Working smoothly  

 

 

 

casters 

Brakes, moving forwards  

 

 

 

Bed rails 

Locking and unlocking 

mechanisms 

 

 

 

Knee raiser  

Hinge point 

 

 

 

Accessories (ex. Trapeze bar, 
handle) 

Mounting, damages, suitability 

 

 

 

Inspection result: 
All values within the allowed range:

    yes  no 

Next inspection date: 

Содержание Bariatric Professional APH188

Страница 1: ...1 Electric Bed Apollo Bariatric Professional APH188 USER S MANUAL This product conforms to the standards of the Medical Directive 93 42 EEC as implemented by Legislative Decree 46 97 and s m e i...

Страница 2: ...2 1 HEAD PANEL 2 BED RAILS 3 Bar 3 HANDCONTROL 4 BED RAILS BUTTON 5 FOOT PANEL 6 WHEELS WITH BRAKE 7 HOUSING FOR LIFTING POLE ON BOTH SIDES 1 6 3 7 4 2 5...

Страница 3: ...safety warnings Pg 13 4 2 Hand control Pg 14 4 2 1 Hand control use Pg 14 4 3 Swivel wheels Pg 15 4 4 Lifting Pole Pg 15 4 5 Bed rails Pg 16 5 CLEANING AND DISINFECTING Pg 17 5 1 General information P...

Страница 4: ...ufacturer in order to avoid any faults This manual provides useful information to the buyer and to the end user on all the functions which will allow a safe and practical daily use of this bed Therefo...

Страница 5: ...is manual The manual may refer to the following groups of people somehow involved with using the bed Owner The owner for example suppliers of medical products dealers health care authorities etc can b...

Страница 6: ...rstand the instructions The safety symbol used does not replace the safety text it refers to It is always necessary to read the relevant safety warning in full and to follow it carefully 2 2 2 Safety...

Страница 7: ...ly disconnect the power plug from the wall clearly mark it as Damaged and deactivate the nursing bed Immediately inform the owner of the bed about this situation In chapter 6 1 you will find a checkli...

Страница 8: ...erification of tests see chapter 6 2 This bed can be used only according to the use conditions described in this manual Any other use is not to be considered as proper 2 3 2 Features Electric height a...

Страница 9: ...nded by 15cm simply by moving the foot panel outwards and the integral mattress support Bed Rails To protect patients from possible accidental falls from the bed the bed is equipped with rails meet cl...

Страница 10: ...s 3 1 Assembly Mounting instructions The assembly of the bed must be carried out by a minimum of two persons Unscrew the 8 hand screws on both the head and foot halves of the frame Use the bars provid...

Страница 11: ...stickers on the mounting points on the underside of the bed frame Equally the green stickers on the foot rest actuator should be matched up with the green stickers on the mounting points The main con...

Страница 12: ...terproof treatment It is possible to use the bed on tiles carpets laminated floors and linoleum A 230 volt power outlet correctly installed must be located as near as possible to the bed If there is t...

Страница 13: ...ssories on the electric bed make sure such accessories do not crush or cut the patient with sharp edges during the legs and backrest adjustments If this condition is not guaranteed the user must not p...

Страница 14: ...Press this button to change the angle of the leg support Backrest and leg support combined adjustment Auto Contour Press this button to change the angle of the backrest and the angle of the leg suppo...

Страница 15: ...e lifting pole must be inserted in the side from where the patient gets in or out of the bed to help him her out Only use the manufacturers lifting pole design for this bed and for specific Bariatric...

Страница 16: ...ositioned at the right distance to allow the rails to snap into place properly Make sure to use only the Apollo original bed rails supplied with the bed When using a bed with bed rails you must always...

Страница 17: ...t not be subjected to a water jet a high pressure cleaner or anything similar Cleaning should be done with a damp cloth If you suspect that water or moisture have entered into the electric components...

Страница 18: ...reading around These recommendations will help to prevent nosocomial infections If the beds are not immediately reused they should be stored away with a cover to protect from dust and accidental conta...

Страница 19: ...important role Do not use alcohol for large surfaces 6 MAINTENANCE Regulations The owners of the nursing beds in accordance with the applicable regulations in force are requested to guarantee the saf...

Страница 20: ...ble damages cable connection Power cord Possible damages cable connection Hand control Possible damages Visual inspection of mechanical components Lifting pole and its housings Possible damages deform...

Страница 21: ...the documentation of the inspection results must be carried out only by experienced personnel who know the proper way and conditions necessary to implement a correct testing The evaluation and the doc...

Страница 22: ...tion Visual inspection of electrical components Identification labels and stickers In place and clearly readable Housing of the control device Properly positioned damages Housing and lifting tubes of...

Страница 23: ...f the first measure ment Current actual value Result bed SK II type B 0 1 ma ma ma Protocol system for testing electrical medical devices Sheet 2 to 2 III Functional control ok not ok Problem descript...

Страница 24: ...s no This inspection protocol included Notes Tested on Examiner Signature Evaluated on Owner staff Signature 6 3 Spare parts Replacement parts are available From Apollo Healthcare Technologies Ltd by...

Страница 25: ...cement Dangerous for your life life threat due to electric shock Before starting any operations remove the plug from the wall outlet Any electrical equipment operations must be carried out only by qua...

Страница 26: ...lution The hand control and or the operating system does not work Power cord not connected Plug not under voltage Hand control plug not properly connected Faulty parts Insert the power cord Check the...

Страница 27: ...rol Type Ti Motion Electric motors for height adjustment of the mattress base and backrest and footrest Type Ti Motion TA7 Series Linear Actuators Max speed at no load full load mm s 38 23 2 8 3 Envir...

Страница 28: ...r and tear or due to excessive exposure to sunlight Damages caused by loss or spillage of liquids Damages from collisions or impacts against walls furniture etc Tears and accidental damages cigarette...

Страница 29: ...ctive 93 42 EEC and to national law no 46 97 and comply with the European Directive on CE marking in addition they are not measuring instruments they are not intended for clinical investigations they...

Отзывы: