CABLE CONNECTORS 

 
Apogee started offering cable connectors on some 
bare-lead sensors in March 2018 to simplify the 
process of removing sensors from weather stations 
for calibration (the entire cable does 

not

 have to be 

removed from the station and shipped with the 
sensor). 
 
The ruggedized M8 connectors are rated IP68, 
made of corrosion-resistant marine-grade stainless-
steel, and designed for extended use in harsh 
environmental conditions.  
  

 

 

Cable connectors are attached directly to the head. 

Instructions 

Pins and Wiring Colors:

  All Apogee connectors 

have six pins, but not all pins are used for every 
sensor. There may also be unused wire colors 
inside the cable. To simplify datalogger 
connection, we remove the unused pigtail lead 
colors at the datalogger end of the cable. 
 
If a replacement cable is required, please contact 
Apogee directly to ensure ordering the proper 
pigtail configuration. 
 

Alignment:

  When reconnecting a sensor, arrows 

on the connector jacket and an aligning notch 
ensure proper orientation.       

 
Disconnection for extended periods: 

 When 

disconnecting the sensor for an extended period 
of time from a station, protect the remaining half 
of the connector still on the station from water 
and dirt with electrical tape or other method. 
 
 

 

A reference notch inside the connector ensures  

proper alignment before tightening. 

 

 

When sending sensors in for calibration, only send the 

sensor head. 

 

Tightening:

  Connectors are designed to be firmly 

finger-tightened only. There is an o-ring inside the 
connector that can be overly compressed if a 
wrench is used. Pay attention to thread alignment 
to avoid cross-threading. When fully tightened, 1-
2 threads may still be visible.  
 

WARNING: 

Do not tighten the connector by 

twisting the black cable or sensor head, only twist 
the metal connector (yellow arrows). 

 

Finger-tighten firmly 

Содержание SQ-627

Страница 1: ...INC 721 WEST 1800 NORTH LOGAN UTAH 84321 USA TEL 435 792 4700 FAX 435 787 8268 WEB APOGEEINSTRUMENTS COM Copyright 2020 Apogee Instruments Inc OWNER S MANUAL EXTENDED RANGE PFD SENSOR Model SQ 627 in...

Страница 2: ...of Compliance 3 Introduction 4 Sensor Models 5 Specifications 5 Deployment and Installation 8 Cable Connectors 8 Operation and Measurement 10 Maintenance and Recalibration 14 Troubleshooting and Cust...

Страница 3: ...of electrical and electronic products with respect to the restriction of hazardous substances Please be advised that based on the information available to us from our raw material suppliers the produc...

Страница 4: ...quantum sensors but measure a wider range of wavelengths can be thought of as an extended range quantum sensor Typical applications of traditional quantum sensors include incoming PPFD measurement ov...

Страница 5: ...respective manuals Model Signal SQ 620 Self powered SQ 622 0 2 5 V SQ 624 4 20 mA SQ 625 0 5 V SQ 626 USB SQ 627 SDI 12 SQ 628 Modbus Sensor model number and serial number are located on the bottom o...

Страница 6: ...onse Time 0 6 s time for detector signal to reach 95 following a step change fastest data transmission rate for SDI 12 circuitry is 1 s Field of View 180 Spectral Range 340 to 1040 nm 5 nm wavelengths...

Страница 7: ...respectively Mean spectral response measurements of six replicate Apogee SQ 600 series Extended Range PFD Sensors Spectral response measurements were made at 10 nm increments across a wavelength range...

Страница 8: ...nted with the cable pointing toward true north in the northern hemisphere or true south in the southern hemisphere Azimuth error is typically less than 0 5 but it is easy to minimize by proper cable o...

Страница 9: ...please contact Apogee directly to ensure ordering the proper pigtail configuration Alignment When reconnecting a sensor arrows on the connector jacket and an aligning notch ensure proper orientation D...

Страница 10: ...tion php version 1 4 August 10 2016 Overview During normal communication the data recorder sends a packet of data to the sensor that consists of an address and a command Then the sensor sends a respon...

Страница 11: ...ommand signals a measurement sequence to be performed Data values generated in response to this command are stored in the sensor s buffer for subsequent collection using D commands Data will be retain...

Страница 12: ...ees if pointed down Available in sensors with serial number 3033 or greater aCC or aCC0 a00101 cr lf Returns mol m 2 s 1 w CRC aCC1 a00101 cr lf Returns millivolt output w CRC aCC2 a00101 cr lf Return...

Страница 13: ...g the sensor version number and xx xx is serial number Running Average Command The running average command can be used to set or query the number of measurements that are averaged together before retu...

Страница 14: ...g may be necessary Dust or organic deposits are best removed using water or window cleaner and a soft cloth or cotton swab Salt deposits should be dissolved with vinegar and removed with a cloth or co...

Страница 15: ...ble Measurement Devices Dataloggers Controllers Meters Any datalogger or meter with SDI 12 functionality that includes the M or C command An example datalogger program for Campbell Scientific datalogg...

Страница 16: ...warranty items to our factory The warranty does not cover equipment that has been damaged due to the following conditions 1 Improper installation or abuse 2 Operation of the instrument outside of its...

Страница 17: ...eeinstruments com to discuss repair or replacement options OTHER TERMS The available remedy of defects under this warranty is for the repair or replacement of the original product and Apogee Instrumen...

Отзывы: