pH Measurement  

7.1

 How to take pH measurements 

Short press            to power on the tester. Rinse the probe in pure water, shake off 

excess water. Insert the probe in your sample solution, make a quick stir and hold 

still. Record the reading when it is stabilized(          appears and stays on screen).   

7.2 

Pure Water pH Measurement 

When testing pure water like tap water, drinking water, RO water and distilled water, 

it will take longer for the readings to get fully stabilized (typically 1-5 minutes). Please be patient. If still not working, add 

Apera 3M KCL (AI1107) to your pure water at the ratio of 1:1000 (e.g. 1 ml KCL to 1000 ml water) to accelerate stabilization 

while minimizing pH change. If the accuracy is not meeting your requirement, please contact us at [email protected] to 

find the specialized meter designed for pure water pH test. 

 

ORP Measurement  

1)  ORP stands for Oxidation-Reduction Potential, measured in mV. 

It’s also called redox. ORP is a measure of the 

cleanliness of water & its ability to break down contaminants. A separate ORP probe (ORP60-DA) needs to be 

installed to be able to measure ORP.   

2)  Power on the tester, unscrew the original probe, and install the ORP60-DA probe, then the tester will automatically 

switch to ORP measurement mode (Refer to Section 15 for how to replace a probe). 

3)  Rinse the probe in distilled water and dry it. Dip the probe in sample solution, shake for a few seconds, and hold still. 

Record the ORP reading after            appears and stays on screen. 

 

 

Probe Cleaning  

1)  The tester is only as accurate as the probe is clean. Always thoroughly rinse off the probe before and after each 

measurement with pure water in a container or with a wash bottle.   

2)  For tough contaminants, detach the sensor shield, soak the probe in Apera probe cleaning solution (AI1166) or 

detergent water for 30 minutes. Then use a soft brush to remove the contaminants. Afterwards, soak the probe in 

3M KCL soaking solution for at least 1 hour. Rinse it off, then re-calibrate the tester before using again. 

 

 

 

 

Содержание ZenTest PH60-Z

Страница 1: ...1 ZenTest PH60 Z Smart pH Tester Kit pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Страница 2: ...2 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attributed to used product Never use the product when it s freezi...

Страница 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Страница 4: ...t press to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting sho...

Страница 5: ...ARDS Preparation before Use 5 1 Pull out the battery insulation slip and take off the probe cap 5 2 Rinse off the probe in pure water preferably distilled or deionized water RO water or tap water is t...

Страница 6: ...0 01 standard pH buffer and repeat Step 6 3 to 6 4 will show up next to and indicating a successful 3 point calibration high low and middle points 6 7 Notes about Calibration 1 The 1st point calibrati...

Страница 7: ...n soak it in 3M KCL 3 5 hours before performing calibration again is pressed before measurement is fully stabilized Wait for to stay on screen before pressing During calibration readings being unstabl...

Страница 8: ...sured in mV It s also called redox ORP is a measure of the cleanliness of water its ability to break down contaminants A separate ORP probe ORP60 DA needs to be installed to be able to measure ORP 2 P...

Страница 9: ...ort press to adjust parameter short press to confirm Short press to exit parameter setting and go back to measurement mode 2 Auto Lock P02 Users can set the auto lock time from 5 to 20 seconds For exa...

Страница 10: ...Factory Default P08 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure a...

Страница 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Страница 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Отзывы: