Apera Instruments ZenTest PH60-Z Скачать руководство пользователя страница 11

 

- 11 - 

 

Troubleshooting G uide  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Trouble

Reason

How to fix

Pressing             too soon 

Wait for          to stay on the screen before pressing 

Incorrect standard solutions 

Reboot tester, calibrate pH 7 first, then pH 4. For details refer to 

Section 5.2 (a)

Poor quality standard solutions 

Replace with fresh and clean standard calibration solutions made 

by legitimate scientific instrument manufacturers.

Contaminated probe 

Use a soft brush to clean the probe with Apera probe cleaning 

solution or detergent water.

Aged probe 

Replace the probe.

Dried-out probe 

Soak in Apera 3M KCL soaking solution for at least 30 minutes.

Probe is not fully submerged

Make sure the probe is immersed in the solution at least 1 inch.

Air bubbles around the sensor shield

Make a quick stir in the solution to remove air bubbles.

Contaminated probe

Use a soft brush to clean the probe with Apera probe cleaning 

solution or detergent water.

Clogged junction

Use a soft brush to clean the probe with Apera probe cleaning 

solution or detergent water, then soak it in Apera 3M KCL soaking 

solution overnight.

Aged probe

Replace the probe.

Testing pure water like 

tap/drinking/RO/distilled water

Be patient, wait for 1-5 minutes to reach a fully stabilized reading. 

If still not stabilizing, add Apera 3M KCL solution to test water at 

1:1000 ratio.

Broken probe

If you don’t find any visible damage of the probe, contact us for 

warranty fulfillment; If there is visible damage, replace the probe.

Instrument defect

Contact us for warranty fulfillment

Probe is not fully submerged in the 

solution 

Make sure the probe is immersed into solution at least 1 inch.

Air bubbles around the sensor shield

Make a quick stir in the solution to remove air bubbles.

Probe is not properly connected or the 

connector is broken. 

Check the probe’s connector, make sure it’s not broken and is 

correctly connected. Align the probe and instrument correctly 

before plugging in. Never force it. Ensure that the probe 

connector is not exposed to the air too long. 

Aged probe

Replace the probe.

Air bubbles around the sensor shield

Make a quick stir in the solution to remove air bubbles.

Clogged junction

Clean the probe with cleaning solution, then soak it in 3M KCL 

storage solution overnight

Comparison with other testers, test strips, 

or drop tests

To compare accuracy with other testers, make sure to perform 

calibration for all testers in the same standard, then test another 

standard. Whichever gives more accurate reading is the more 

accurate one. Test strips or drop tests’ accuracy is not comparable 

to pH meters’.

Poor quality standard solutions

Replace with fresh and clean standard calibration solutions made 

by legitimate scientific instrument manufacturers.

The probe is not suitable for your 

appliacation.

Contact us to find the most appropriate product for your specific 

application.

Cannot calibrate

Reading is always slowly 

changing, won’t stabilize.

Display similar readings in 

any solutions or always 

display 7.0 pH

Reading keeps jumping

Calibration is successful, 

but reading is not accurate

Содержание ZenTest PH60-Z

Страница 1: ...1 ZenTest PH60 Z Smart pH Tester Kit pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Страница 2: ...2 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attributed to used product Never use the product when it s freezi...

Страница 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Страница 4: ...t press to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting sho...

Страница 5: ...ARDS Preparation before Use 5 1 Pull out the battery insulation slip and take off the probe cap 5 2 Rinse off the probe in pure water preferably distilled or deionized water RO water or tap water is t...

Страница 6: ...0 01 standard pH buffer and repeat Step 6 3 to 6 4 will show up next to and indicating a successful 3 point calibration high low and middle points 6 7 Notes about Calibration 1 The 1st point calibrati...

Страница 7: ...n soak it in 3M KCL 3 5 hours before performing calibration again is pressed before measurement is fully stabilized Wait for to stay on screen before pressing During calibration readings being unstabl...

Страница 8: ...sured in mV It s also called redox ORP is a measure of the cleanliness of water its ability to break down contaminants A separate ORP probe ORP60 DA needs to be installed to be able to measure ORP 2 P...

Страница 9: ...ort press to adjust parameter short press to confirm Short press to exit parameter setting and go back to measurement mode 2 Auto Lock P02 Users can set the auto lock time from 5 to 20 seconds For exa...

Страница 10: ...Factory Default P08 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure a...

Страница 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Страница 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Отзывы: