- 10 - 

4) 

Auto. Power off (P04)

 

  The auto. power off time can be set to 10 to 20 minutes. For example, if 15 minutes is 

set, the meter will automatically shut down after 15 minutes if no operation; when 

“Off” is set, the auto. power off 

function will be turned off. Long press          to manually shut down the meter. 

5) 

pH Calibration Reminder (P07) 

– set X hours (H) Or X days (D) in ZenTest mobile app – settings – Parameter – 

pH 

– Calibration Reminder. On the meter, you can only check the values that’s been set up on ZenTest App. For 

example, if 3 days is set up, the Er6 icon (see Figure-4) will appear in the lower right corner of the LCD screen in 3 

days to remind you to perform calibration, also in the ZenTest App there will be a pop-up reminder. After calibration 

is finished or the reminder setting is cancelled in the ZenTest App, the Er6 icon will disappear. 

6) 

pH Back to Factory Default (P08) 

– Select “Yes” to recover instrument calibration to theoretical value. This 

function can be used when instrument does not work well in calibration or measurement. Calibrate and measure 

again after setting the instrument back to factory default.   

 

Technical Specifications  

pH 

Range 

-2.00 to 16.00 pH 

Resolution 

0.01 pH 

Accuracy 

±0.01 pH ±1 digit 

Calibration Points 

1 to 3 points 

Auto. Temperature Compensation 

– 50˚C (32 – 122˚F) 

ORP (mV) 

Range 

-1000 mV to 1000 mV 

Accuracy 

±0.2% F.S 

Temperature 

Range 

0 to 50˚C (32-122˚F) 

Accuracy 

±0.5˚C 

 

Icons and  Functions  

 

 

 

 

 

 

 

Calibrated points 

 

Self-Diagnosis Symbol 

Er1, Er2, Er3, Er4,Er5, Er6 

Stable reading indicator 

 

Waterproof Rating 

IP67, floats on water 

Reading Lock 

HOLD 

Power 

DC3V, AAA batteries*4 

Bluetooth Signal 

 

Battery Life 

200 Hours 

Low power reminder 

           

Backlight 

White: Measurement

Green: Calibration

Red: Alarm 

Auto. Power Off 

Automatically power off if no operation for 10 minutes

 

Dimension/Weight 

Instrument

40×40×178mm/133g; case

255×210×50mm/680g; 

LCD Display 

pH calibration reminder 

 

pH alarm triggered 

 

Содержание ZenTest PH60-Z

Страница 1: ...1 ZenTest PH60 Z Smart pH Tester Kit pH Temp ORP User Manual APERA INSTRUMENTS LLC aperainst com V1 5...

Страница 2: ...2 ATTENTION Water droplets are added during production to maintain the moisture of the probe This is normal practice and should not be attributed to used product Never use the product when it s freezi...

Страница 3: ...erate the tester without connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LED display 1 Basic Mode digital display cal...

Страница 4: ...t press to turn on off backlight 1 In measurement mode long press to turn on off Bluetooth receiver When turned on will be flashing when connected to smartphone will stay on 3 In parameter setting sho...

Страница 5: ...ARDS Preparation before Use 5 1 Pull out the battery insulation slip and take off the probe cap 5 2 Rinse off the probe in pure water preferably distilled or deionized water RO water or tap water is t...

Страница 6: ...0 01 standard pH buffer and repeat Step 6 3 to 6 4 will show up next to and indicating a successful 3 point calibration high low and middle points 6 7 Notes about Calibration 1 The 1st point calibrati...

Страница 7: ...n soak it in 3M KCL 3 5 hours before performing calibration again is pressed before measurement is fully stabilized Wait for to stay on screen before pressing During calibration readings being unstabl...

Страница 8: ...sured in mV It s also called redox ORP is a measure of the cleanliness of water its ability to break down contaminants A separate ORP probe ORP60 DA needs to be installed to be able to measure ORP 2 P...

Страница 9: ...ort press to adjust parameter short press to confirm Short press to exit parameter setting and go back to measurement mode 2 Auto Lock P02 Users can set the auto lock time from 5 to 20 seconds For exa...

Страница 10: ...Factory Default P08 Select Yes to recover instrument calibration to theoretical value This function can be used when instrument does not work well in calibration or measurement Calibrate and measure a...

Страница 11: ...ble damage replace the probe Instrument defect Contact us for warranty fulfillment Probe is not fully submerged in the solution Make sure the probe is immersed into solution at least 1 inch Air bubble...

Страница 12: ...ged product attributable to responsibility of APERA INSTRUMENTS LLC for a period of TWO YEARS SIX MONTHS for the probe from the delivery This limited warranty does NOT cover any damages due to Acciden...

Отзывы: