Symbol 

Self-Diagnosis information 

Potential problems and how to fix 

 

 

 
 

The pH calibration 
solution cannot be 
recognized by the meter. 

1. 

Make sure the probe is fully immersed in the 
calibration solution.

 

2. 

Check if calibration solution is expired or polluted.

 

3. 

1st point of pH calibration must be pH 7.00 or 6.86. 
See 6.2 (1).

 

4. 

Please check whether pH electrode is damaged or 
broken. If so, please replace with a new one.

 

5. 

The glass bulb or junction is severely 
contaminated. Please use a soft brush with soap 
water to clean it thoroughly. Then soak it in 3M KCL 
3-5 hours before performing calibration again.

 

 

 

 

Is pressed before 

measurement is fully 
stable   

Wait for 

  to come up and stay on screen 

before pressing 

       

 

During calibration, 
readings being unstable for 
over 3 minutes 

1. 

Please check whether pH electrode is damaged or 
broken. If so, please replace with a new one.

 

2. 

The glass bulb or junction is severely 
contaminated. Please use a soft brush with soap 
water to clean it thoroughly. Then soak it in 3M KCL 
overnight before performing calibration again.

 

3. 

The electrode is aged (used for over a year and 
has a much slower response). A replacement is 
needed.

 

 

pH electrode zero electric 
potential out of range   
(<-60mV or >60mV) 

1. 

Check whether pH buffer solutions comply with the 
USA or NIST standard.

 

2. 

Check whether pH buffers are expired or 
contaminated.

 

3. 

Please check whether pH electrode is damaged or 
broken. If so, please replace with a new one.

 

4. 

The electrode is aged (used for over a year and 
has a much slower response). A replacement is 
needed.

 

5. 

The electrode is invalidated (Er4/Er5 repetitively 
appears, and problems 1, 2, 3 are excluded). A 
replacement is needed.

 

pH electrode slope out of 
range (<85% or >110%)

 

 

The calibration reminder is 
triggered. I

t’s t

ime to 

perform a new pH 
calibration 

Perform pH calibration or cancel calibration reminder in 
ZenTest App settings.

 

7.  pH Measurement 

  PH60S-Z Spear pH tester is designed for direct spear test of solid samplings such as cheese, 

fruit, sushi rice, meat, dough. It also works well in general water solutions. 

  When performing spear tests in solid samples, pay attention to insert probe evenly, and be 

careful to prevent probe from damage. If the medium is too hard, please bore a small hole with 

a clean knife before inserting the spear probe. 

Содержание AI3711

Страница 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Страница 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Страница 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Страница 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Страница 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Страница 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Страница 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Страница 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Страница 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Страница 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Страница 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Отзывы: