background image

 

- 6 -

 

is stabilized (

stays on the LCD screen), short press 

  to complete 1

st

 point calibration. 

After calibration is completed, the tester will return to measurement mode. Icon 

  will appear 

at the bottom left of the LCD screen. indicating a successful 1

st

 point pH calibration. 

  To calibrate 2

nd

 point, use 4.00 pH buffer and repeat Step 6.3 to 6.4 (Do NOT turn off the tester 

after you finish pH 7 calibration). 

  will display next to 

, indicating a successful 2-point 

pH calibration (low and middle points).   

  If necessary (target pH>8.00), calibrate 3

rd

 point using 10.01 standard pH buffer and repeat 

Step 6.3 to 6.4, 

  will show up next to 

  and 

, indicating a successful 3-point 

calibration (high, low, and middle points). 

  Notes about Calibrationhe   

1)  The 1st point calibration must be 7.00 pH. Perform the 2nd and 3rd point calibrations (4.00, 

10.01, 1.68, or 12.45) immediately after the 1st point calibration is finished. Do NOT turn off the 

meter before you calibrate 2nd or 3rd point. Otherwise, you will need to restart the calibration 

process with 7.00 pH first.   

2)  The pH 4.00 and 7.00 buffer solutions poured into the calibration vials can be used for up to 10 

times as long as they are not contaminated and the bottles are capped when not in use. pH 

10.01 can only be used for up to 5 times as it will lose its accuracy much faster. After that, 

replace the buffer solutions in the calibration vials with new ones to keep the accuracy. Keeping 

the freshness and cleanliness of calibration buffers is essential for accurate pH measurement. 

3)  The tester can perform 1 to 3 points of automatic calibration and can recognize 5 types of pH 

standard solutions. For details, please refer to the following table: 

 

Calibration

 

USA Series

 

NIST Series

 

Indication 

icon

 

Recommended

 

1-point   

 

7.00 pH

 

6.86 pH

 

 

Accuracy 

requirement ≥ 0.1 

pH

 

2-pt

 

Option 

A

 

1

st

 pt: 7.00 pH

 

2

nd

 pt: 4.00 pH or 

1.68 pH

 

Option 

A

 

1

st

 pt: 6.86 pH

 

2

nd

 pt: 4.01 pH or 

1.68 pH

 

 

Range < 7.00 pH

 

Option 

B

 

1

st

 pt: 7.00 pH

 

2

nd

 pt: 10.01 pH or 

12.45 pH

 

Option 

B

 

1

st

 pt: 6.86 pH

 

2

nd

 pt: 9.18 pH or 

12.45 pH

 

 

Range >7.00 pH

 

3-pt

 

1

st

 pt: 7.00 pH

 

2

nd

 pt: 4.00 or 1.68 pH

 

3

rd

 pt: 10.01 or 12.45 pH

 

1

st

 pt: 6.86 pH

 

2

nd

 pt: 4.01 or 1.68 pH

 

3

rd

 pt: 9.18 pH or 12.45 

pH

 

 

 

Range: 0 to 14.00 

pH

 

 
 

  For the self-diagnosis information, please refer to the table below: 

 

 

Содержание AI3711

Страница 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Страница 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Страница 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Страница 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Страница 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Страница 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Страница 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Страница 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Страница 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Страница 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Страница 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Отзывы: