14. Probe Replacement 

pH probes don’t last forever. Every probe will eventually age and fail even if you don't use it that often. 
The typical service life of Apera probes is 12-24 months depending on the frequency of usage and 
how well you keep it clean and properly stored. We recommend replacing your probe every 1 to 2 
years to ensure the best performance

 

 

To replace a probe: 

 

1) Take off the probe cap; 2) Screw off the probe ring 3) Unplug the 
probe;  4)  Plug  in  the  new  replacement  probe  (pay  attention  to  the 
probe’s position); 5) Screw on the probe ring tightly. Soak the probe in 
3M KCL for 5-15 minutes. Then perform calibration before testing.

 

 

The replacement probes that are compatible with PH60S-Z are:

 

 

PH60S-DE (Spear pH probe for solids/semi-solids pH testing), 
PH60-DE (Regular pH glass bulb probe), PH60F-DE (Flat pH 
probe for surface pH testing), 

 

 

ORP60-DA (ORP Probe).

 

15. Troubleshooting Guide 

Trouble 

Reason 

How to fix 

Cannot 

calibrate 

Pressing 

  too soon 

Wait for 

  to stay on screen before pressing 

 

Incorrect standard 

solutions 

Reboot tester, calibrate pH 7 first, then pH 4. For details 

refer to Section 5.2 (a).

 

Poor quality standard 

solutions 

Replace with fresh and clean standard calibration 

solutions made by legitimate scientific instrument 

manufacturers. 

Contaminated probe 

Clean the probe with Apera cleaning solution or 

detergent water. 

Aged probe 

Replace the probe. 

Dried-out probe 

Soak the probe in 3M KCL solution for at least 15 

minutes. 

Probe is not fully 

submerged in the solution 

Make sure the probe is fully immersed in the 

solution at least 1 inch deep. 

Air bubbles around the 

sensor 

Make a quick stir in the solution to remove air 

bubbles. 

Reading is 

always slowly 

changing, won’t 

stabilize 

Contaminated probe 

Clean the probe with 

Aprea’s cleaning solution or 

detergent water. 

Clogged junction 

Clean the probe with Aprea’s cleaning solution, then 

soak it in 3M KCL soaking solution overnight. 

Aged probe 

Replace the probe. 

Testing pH of low iconic 

strength solutions like tap 
water, drinking water, RO 

water 

Be patient, wait for 1-5 minutes to reach a fully 

stabilized reading. If still not stabilizing, add 1ml of 

3M KCL solution to 1000ml of test solution (or 1 

teaspoon to 1 gallon). 

Содержание AI3711

Страница 1: ...1 ZenTest PH60S Z Smart Spear pH Tester pH Temp ORP Instruction Manual APERA INSTRUMENTS Europe GmbH www aperainst de...

Страница 2: ...Calibration 5 7 pH Measurement 7 8 ORP Measurement 8 9 Probe Cleaning 8 10 Probe Storage 8 11 Parameter Setting 9 12 Technical Specifications 10 13 Icons and Functions 10 14 Probe Replacement 11 15 Tr...

Страница 3: ...connecting to a smartphone Table 1 Funtions on 60 Z Tester and ZenTest Mobile App Functions 60 Z Tester ZenTest Mobile App Display LCD display 1 Basic Mode digital display calibration info Swipe to sw...

Страница 4: ...on Please install batteries according to the following steps Please note direction of batteries All POSITIVE SIDES FACING UP Wrong installation of batteries will cause damage to the tester and potenti...

Страница 5: ...calibration bottles to about half volume of the bottle Long press to enter calibration mode screen turns green Insert the probe into pH 7 00 buffer solution make a quick stir and hold still When the r...

Страница 6: ...H first 2 The pH 4 00 and 7 00 buffer solutions poured into the calibration vials can be used for up to 10 times as long as they are not contaminated and the bottles are capped when not in use pH 10 0...

Страница 7: ...ibration again 3 The electrode is aged used for over a year and has a much slower response A replacement is needed pH electrode zero electric potential out of range 60mV or 60mV 1 Check whether pH buf...

Страница 8: ...d to be able to measure ORP 2 Power on the PC60 Z tester unscrew the original probe and install the ORP60 DA probe then the tester will automatically switch to ORP measurement mode Refer to Section 14...

Страница 9: ...rt press to manually turn the backlight on or off 4 Auto Power off P04 The auto power off time can be set to 10 to 20 minutes For example if 15 minutes is set the meter will automatically shut down af...

Страница 10: ...ature Compensation 0 50 C 32 122 F ORP mV Range 1000 mV to 1000 mV Accuracy 0 2 F S Temperature Range 0 to 50 C 32 122 F Accuracy 0 5 C Calibrated points Self Diagnosis Symbol Er1 Er2 Er3 Er4 Er5 Er6...

Страница 11: ...ng Incorrect standard solutions Reboot tester calibrate pH 7 first then pH 4 For details refer to Section 5 2 a Poor quality standard solutions Replace with fresh and clean standard calibration soluti...

Страница 12: ...ean the probe with cleaning solution then soak it in 3M KCL soaking solution overnight Comparison with other testers test strips or drop tests To compare with other testers make sure to perform a 2 po...

Отзывы: