background image

19

 

mixed). If this takes place, the sensors must be switched off (by setting 

ID

=

16

) and deleted from the recorder 

list (see item 2 below), and then the indexing process should be restarted. 

b)

 

programmable (using the ARSOFT-CFG software and, optionally, the keypad of the AR407 recorder) 
Using the ARSOFT-CFG software, one must assign different indices (the 

ID

 parameter to the measurement 

sensors and then the same values should be used in the target measurement channels of the AR407 recorder 
in the 

Address of assigned sensor (ID)

 parameter (in the 

Measurement channels 

configuration menu, chapter 

12.4). 
In order to present two measurement channels in the case of two-channel sensors (AR433/434/435/436/437), 
the same 

ID

 number should be assigned to two measurement channels of the AR407 recorder.  

 

NOTE:  

!

 

Each radio sensor working in the system 

must

 have a different identification number (the 

ID

 parameter)! 

 

2.

  

In order to delete radio sensors from the measurement

 list of the AR407 recorder, one must only set, for 

the selected channel, the 

Address of assigned sensor (ID)

 or 

Assigned sensor type

 parameter to the value 

None

 (

Measurement channels 

configuration menu, chapter 12.4). In order to quickly delete the entire 

measurement list (all assigned sensors), one can use the 

Delete list of radio sensors

 function available in the 

following location: 

Main Menu

 -> 

Memory and file option

 -> 

Reset to default settings

, chapter 12.3. 

 

Moreover, sensors that are not in use should be switched off

 in order to save the battery and to eliminate 

their influence on the system in operation. To do so, the 

ID

 parameter of the sensor must be set to the value of 

16

; this can be done in two ways:  

a)

 

with the power supply of the AR407 recorder 

switched off

 or with the recorder outside of the radio 

range, press the 

ADD/REF/OFF

 button in the sensor for about 10 sec.; the sensor will be switched off 

after no more than 35 sec.; this can be done simultaneously with several sensors 

b)

 

using the ARSOFT-CFG software and the connected USB cable 

 

NOTE:  

!

 

If radio sensors are not in use, it is recommended to completely switch off the radio scanning in the AR407 by 
setting in the unused measurement channels the 

Assigned sensor type

  to the 

None (channel switched off)

 value. 

 

In order to test the range in the eventual place of installation, the functions provided by the 

ADD/REF/OFF

 button 

in the AR43x sensor can be used (using the methods described in the user’s manual of the sensor, chapter 5.1). 

For optimum performance of the system and in the event of communication problems

 (between the sensors 

and the base station of the AR407)

, read the operating notes in chapter 16.

 

 
 
 

12.6. CONFIGURATION OF ALARM OUTPUTS

 

The 

Outputs (alarms) configuration

 menu comprises four identical sets of parameters grouped according to the 

output number described in the Table 12.6.1 and other items listed in Table 12.6.2.  
For alarm states, it is possible to program the value of the output signal (the filling coefficient with the impulse 
period set), which can be useful for signaling purposes or for adjustment of power. 

Table 12.6.1. Configuration parameters in the 

Outputs (alarms) configuration

 menu for the selected alarm output 

Parameter  

 

Range of variability of the parameter and description 

Default 

settings 

 

Assigned inputs

 

numbers of the active measurement channels assigned to the selected alarm; 
the assignment is made in the 

Measurement channel 

configuration menu 

(chapter 12.4, parameter 

Assignment of alarm outputs

1÷16

 or 

None

 

Содержание AR407

Страница 1: ...USER INSTRUCTION AR407 16 CHANNEL RADIO AND WIRED DATA RECORDER Version 1 0 0 2017 04 10 93 87 10310 Tel 02 932 2326 Fax 02 903 0080 Ext 6867 Email info omi co th Website www omi co th...

Страница 2: ...10 FUNCTIONS OF HARDWARE AND SCREEN BUTTONS 9 11 DESCRIPTION OF DATA SHOWN ON THE LCD DISPLAY 11 11 1 UPPER AND LOWER STATUS BAR 11 11 2 VIEW OF A MEASUREMENT GROUP TEXT ONLY 12 11 3 VIEW OF A MEASUR...

Страница 3: ...e device d ensure proper operating conditions compliant with the technical specification of the device chapter 5 power supply voltage humidity temperature etc do not expose the device to direct and in...

Страница 4: ...fox etc the site contains information about active measurement channels time status of the outputs recording etc with the possibility to show diagrams using the Google Chart API service diagrams requi...

Страница 5: ...ltaneous recording of data from all active measurement channels an enclosure for panel installation leak tightness from the front side IP30 compliance with the R TTE 99 5 EC Directive possible update...

Страница 6: ...ual COM serial port MODBUS RTU protocol master mode host support of USB memory pendrive up to 4 GB writing speed approx 135kB s depending on the memory RS485 galvanic separation MODBUS RTU protocol Ma...

Страница 7: ...mbering and description of the connections on the back panel and method of connecting probes and electrical signals Terminal clamps connections Description ANTENNA SMA socket to connect an antenna for...

Страница 8: ...OMx serial port x port number 1 2 etc The serial port uses the MODBUS RTU protocol In the internal memory two configuration text files are shown AR407 cfg and AR407 txt chapter 12 Communication with t...

Страница 9: ...ord protection in the Access and other settings menu is switched on then the password must be entered chapter 12 9 closes the message window appearing on the screen chapter 15 UP or DOWN screen button...

Страница 10: ...epending on the value of the F button function parameter in the Access and other settings menu chapter 12 9 Status signa ling icon Device status device status screen factory setting function available...

Страница 11: ...g of the individual graphic elements is described below a lower status bar Fig 11 1 2 Lower status bar Object Description and marking in the contents of the instructions 6 no USB connection with the c...

Страница 12: ...VIEW OF A MEASUREMENT GROUP TEXT ONLY Fig 11 2 Appearance of a measurement group screen in the TEXT ONLY mode presentation max 16 channels is possible 11 3 VIEW OF A MEASUREMENT GROUP TEXT AND A BAR...

Страница 13: ...er configuration can be performed using one of the following three methods do not use them at the same time 1 From the film keypad and a touch screen located on the front panel of the device from the...

Страница 14: ...e port and the MODBUS Address of the device in the case of the RS485 make sure that the serial port drivers have been installed correctly chapter 8 disconnect for a few seconds and then reconnect the...

Страница 15: ...dic daily hourly recording is active within the range of the Time defined by the parameters Start and End of time limit Above the permission threshold recording is active when the measured value of th...

Страница 16: ...rnal only archive files are created only in the internal memory Internal only Auto select archive files are created in the USB memory when detected or the internal memory when there is no USB memory A...

Страница 17: ...e selected channel 1 16 Parameter Range of variability of the parameter and description Default settings Name unit and group edition of the name of the channel and the group max length 16 characters a...

Страница 18: ...in its user s instruction and in the Communication settings chapter 12 8 set the RS485 operation mode parameters to Master and the Baud rate for RS485 to one that is compatible with the device If the...

Страница 19: ...n be done in two ways a with the power supply of the AR407 recorder switched off or with the recorder outside of the radio range press the ADD REF OFF button in the sensor for about 10 sec the sensor...

Страница 20: ...r 9999 19999 units 2 50 0 C Alarm hysteresis 1 0 500 0 C or 0 5000 units 2 0 0 C Output signal value duty cycle 1 100 1 step output control value for the activated alarm filling coefficient with the O...

Страница 21: ...above in chapter 12 6 in Table 12 6 2 in item Email settings and status In order to eventually start the configured service the Email alarm notifications parameter must additionally be set to the Enab...

Страница 22: ...IP number and router modem configuration To facilitate access to the networks with varying public IP address one can start the built in DDNS server service described in chapter 12 8 1 The selection o...

Страница 23: ...network 192 168 0 10 DNS server 0 0 0 0 255 255 255 255 domain name server address DNS 192 168 0 10 Dynamic DNS server client DDNS Disabled A DDNS service that enables access over the Internet to a re...

Страница 24: ...manent nature which indicates inadequate Internet connection improper configuration of the connection or service codes 2 to 12 e g DDNS 5 means invalid username or password DDNS 8 means invalid host n...

Страница 25: ...ffices for at least 5 years of continuous operation Table 12 10 Configuration parameters in the Time and date menu Parameter Range of variability of the parameter Date yyyy mm dd 2008 06 01 2099 12 31...

Страница 26: ...fg file off line NEW USER parameter configuration from the keypad and touch screen user level NEW CH_TEXT name configuration by way of modification of the AR407 txt file creation of a new csv file 4 I...

Страница 27: ...s furniture etc b change the radio channel in all devices in the system in each separately to a different one c check the battery level in the AR43x sensor and if necessary replace the battery with a...

Страница 28: ...line Fig 17 at the start of the line 2 x 820 to the ground and 5 V of the MASTER and 150 between lines at the end of the line 150 between lines termination and polarizing resistors when the MASTER is...

Страница 29: ...0x84 0x02 0x0001 0x5130 19 MODBUS TCP SERIAL TRANSMISSION PROTOCOL The MODBUS TCP protocol is accessible to the Ethernet RJ45 interface and uses the TCP IP transport layer The parameters used by the s...

Страница 30: ...s HB and hours LB R W 0x22 34 0x0000 0x3B3B minutes HB and seconds LB R W 0x23 35 0x0101 0x630C years HB and months LB Parameter End of the time limit chapter 12 2 R W 0x24 36 0x0100 0x1F17 days HB an...

Страница 31: ...7 R W 0xFB 251 60 14400 Parameter Radio scanning interval chapter 12 8 number of seconds s R W 0xFC 252 0 6 Parameter Radio channel number chapter 12 8 R W 0xFD 253 0 1 Parameter USB operation mode ch...

Страница 32: ...32 21 USER S NOTES...

Отзывы: