12. DENTAL UNIT MAINTENANCE SCHEDULE
WHEN
PART
ACTION
REFERENCE
PARAGRAPH
Daily activity
start.
Condensate drain cock.
Eliminate condensate from air ducts.
See paragraph 9.2.
Spray ducts
Duct 2-minute flushing
a)
Manual: activate each instrument with no load
b)
Automatic: activate the LONG FLUSHING cycle
See paragraph 7.5
Air-liquid separator canister by
CATTANI
Insert inside each suction filter a defoaming tablet.
See paragraph 9.5.
Disposable infection control sheaths.
Apply flexible sheaths and disposable barriers on
dental chair and dental unit.
/
Instruments.
Lubricate turbine and micromotor handpieces.
See documentation
enclosed with instrument.
W.H.E. system
(if the dental unit was left off for more
than 7 days)
Empty the W.H.E. system water circuit
See paragraph 7.3.
After each
patient.
Spray ducts.
20-second flushing of the instruments used on the
patient (Manual or QUICK FLUSHING)
See paragraph 7.5.
Instruments.
Disinfect outside.
See documentation
enclosed with instrument.
Camera.
Disinfect outside.
See paragraph 5.8.
Curing Light.
Disinfect outside.
See paragraph 5.7.
Suction tubes.
Carry out a STANDARD flushing cycle (if any) or suck
in about half a litre of 6% diluted
STER 3 PLUS
solution
with each of the suction tubes used.
See paragraph 9.4.
Flexible protections.
Sterilise the contaminated flexible protections.
/
Disposable barriers.
Replace disposable barriers.
/
Contaminated surfaces.
Clean surfaces with
STER 1 PLUS
.
/
Daily activity
end.
Spray ducts.
Disinfection of spray water ducts with manual cycle (if
the S.H.S. system is fitted).
See paragraph 7.2.1.
Disinfection of spray water ducts with semi-automatic
cycle (if the BIOSTER S system is fitted).
See paragraph 7.2.
Bowl filter.
Clean filter in running water.
The content must be disposed of separately.
See paragraph 7.1.
Bowl unit.
Clean with off-the-shelf detergents formulated for
ceramic materials.
Do not use acids or abrasive products.
See paragraph 7.1.
Suction filter.
Check filter and, in case suction capacity is reduced,
change it (code 97461845)
See paragraph 9.3.
Suction tubes.
Carry out an automatic flushing cycle (if any) or suck in
about half a litre of 6% diluted
STER 3 PLUS
solution
with each of the suction tubes used.
See paragraph 9.4.
Hydraulic saliva ejector.
Clean the filter of the saliva ejector terminal.
See paragraph 6.4.
Flexible protections.
Sterilise the present flexible protections.
/
Disposable barriers.
Eliminate all used disposable barriers.
/
Dental unit and chair surfaces.
Clean surfaces with
STER 1 PLUS
.
/
Содержание A5 CONTINENTAL
Страница 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Страница 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Страница 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Страница 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Страница 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Страница 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Страница 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Страница 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Страница 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Страница 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Страница 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Страница 99: ......
Страница 100: ...www cefla com...