7.5. AUTOMATIC INSTRUMENT FLUSHING CYCLE
Description of the system.
This system allows carrying out an automatic FLUSHING cycle to renew
water present in the water ducts of the instruments on the dentist’s and
the assistant’s boards and the water-to-cup duct.
The flushing cycle can be carried out with mains water, with W.H.E.
system (if installed) or with S.H.S. system (if installed).
Available cycles are two:
-
QUICK FLUSHING (cycle time: 20 seconds).
-
LONG FLUSHING (cycle time can be set from 2 to 10 minutes).
It is advisable to carry out a LONG FLUSHING cycle at the
beginning of each working day and a QUICK FLUSHING
cycle between two patients.
Flushing cycle setting.
• If the S.H.S. system is installed and you wish to carry out the flushing
cycle with distilled water, make sure that the relevant icon on the console
DISPLAY is on (see paragraph 7.2.).
It is advisable to execute the flushing cycle with tank
completely full.
• Use the DISPLAY to access the "FLUSHING cycle setting" menu, and
set cycle time (see paragraph 5.1.1.1).
• Position the special container (
d
) for the instruments to be treated on
the bowl.
For the syringe instrument, it is necessary to remove the
handle and completely insert the dedicated adapter (f)
(second click). The heating system must be switched off.
Turbine and scaler hoses must be inserted without the
handpiece.
• Insert the special supplied container (
e
) under the cup spout.
• Make sure that spray cocks (
g
), positioned on board lower side, are fully
open.
Performing the flushing cycle.
• Touch "
PLAY
” icon button (
A
) on the LCD Touch DISPLAY to start the
flushing cycle (see paragraph 5.1.1.1.).
• At the end of the flushing cycle, put the instruments extracted back into
place to return to the working condition.
Flushing cycle interruption.
• Touch “
PAUSE
” icon button (
A1
) to interrupt the flushing cycle at any
time.
• Touch the "
STOP
” icon button (
A2
) within 5 seconds to stop the flushing
cycle in progress and cancel it definitively.
• After 5 seconds, touch “
PAUSE
” icon button (
A1
) to resume flushing
cycle from where it was interrupted.
Warning messages on DISPLAY.
If system detects a malfunction, a warning message will be displayed (see
paragraph 10).
Содержание A5 CONTINENTAL
Страница 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Страница 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Страница 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Страница 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Страница 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Страница 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Страница 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Страница 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Страница 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Страница 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Страница 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Страница 99: ......
Страница 100: ...www cefla com...