A5 - A6 PLUS - L6
EN
43
Removable hose.
The micromotor features a removable hose to make cleaning operations easier (see paragraph 5.).
Cleaning and maintenance.
Refer to the specific instructions provided with the instrument.
It is recommended to use Daily Oil Plus (CEFLA s.c.) for lubrication.
•
Do not soak the instrument in disinfectant fluids or detergents.
•
Inadvisable products: abrasive products and/or products containing acetone, chlorine and sodium hypochlorite.
Sterilization.
Instrument handpiece only: see paragraph 1.6.
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
Safety standards
•
The instrument is supplied NOT sterile and must be sterilized before use.
Carefully read the operating instructions supplied with the instrument before attempting to sterilize.
•
Never put the contra angle on the micromotor while it is running.
•
The drill release button must not be pressed during operation!
Friction between the button and micromotor rotor overheats the head and may cause burns.
•
The patient’s internal tissues (tongue, cheeks, lips, etc.) must be protected against contact with the button by using suitable
instruments (mirrors, etc.).
•
The drills and various instruments attached to the handpieces must comply with Biocompatibility Standard ISO 10993.
5.5.1.
RESTORATIVE OPERATING MODE
Characteristics.
- Speed range: 100 to 40,000 Rpm (1:1 handpiece).
- Reduction ratio selection.
- Display of the instantaneous maximum speed.
- Direct selection of the set working programmes.
Menu with micromotor extracted but not active.
All icon buttons are active and all available functions can be edited (see
paragraph 5.5.).
Any edited setting or value will be automatically saved inside the
selected working programme (e.g. P1).
Menu with micromotor extracted and active.
The following functions can be changed:
• Adjust the maximum rotation speed of the drill (
E
) using the buttons “
+
”
or “
-
”“.
5.5.2.
ENDODONTIC OPERATING MODE
Characteristics.
- Speed value adjustable from 100 to 1200 Rpm with value always
referring to the drill, based on the set reduction ratio,
- Torque value adjustable from 0.1 to 5.0 Ncm, 1:1 reducer excluded (4.5
Ncm),
- Reduction ratio selection with display of certified contra angles,
- Drill rotation set at maximum torque.
- Display of the instantaneous maximum speed.
- Direct selection of the set working programmes.
Menu with micromotor extracted but not active.
In addition to the common buttons, the following buttons are also available
in ENDODONTIC mode:
B
Torque value adjustment
S
Drill rotation at maximum torque.
Содержание A5 CONTINENTAL
Страница 1: ...97055066 Rev 00 2019 03 OPERATOR S MANUAL A5 A6 PLUS L6 EN...
Страница 87: ...A5 A6 PLUS L6 EN 87 11 1 A5 CONTINENTAL DIMENSIONAL CHARACTERISTICS...
Страница 88: ...A5 A6 PLUS L6 88 EN 11 2 A5 INTERNATIONAL DIMENSIONAL CHARACTERISTICS...
Страница 89: ...A5 A6 PLUS L6 EN 89 11 3 DIMENSIONAL CHARACTERISTICS OF A5 CART MODEL...
Страница 90: ...A5 A6 PLUS L6 90 EN 11 4 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE CART MODEL...
Страница 91: ...A5 A6 PLUS L6 EN 91 11 5 DIMENSIONAL CHARACTERISTICS OF A5 SINGLE WATER UNIT MODEL...
Страница 92: ...A5 A6 PLUS L6 92 EN 11 6 A6 PLUS CONTINENTAL MODEL DIMENSIONAL CHARACTERISTICS...
Страница 93: ...A5 A6 PLUS L6 EN 93 11 7 A6 PLUS INTERNATIONAL MODEL DIMENSIONAL CHARACTERISTICS...
Страница 94: ...A5 A6 PLUS L6 94 EN 11 8 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL...
Страница 95: ...A5 A6 PLUS L6 EN 95 11 9 DIMENSIONAL CHARACTERISTICS OF L6 CONTINENTAL MODEL in HYBRID version...
Страница 96: ...A5 A6 PLUS L6 96 EN 11 10 DIMENSIONAL CHARACTERISTICS OF L6 SIDE DELIVERY MODEL...
Страница 99: ......
Страница 100: ...www cefla com...