2 - 49
Chapter 2 Operation
Error Message
Description
RANGE
ERROR:F1 > F2
The start (F1) frequency is greater than the
stop (F2) frequency.
RANGE
ERROR:D1 > D2
The start (D1) distance is greater than the
stop (D2) distance.
RANGE
ERROR:D2 >
DMax=xx.x ft (m)
The stop distance (D2) exceeds the maxi-
mum unaliased range. This range is deter-
mined by the frequency span, number of
points, and relative propagation velocity:
Maximum Unaliased Range
V
F
F
f
=
´
-
( .
) (
) ( )
1 5 10
129
2
1
8
Where: F1 is start frequency
F2 is stop frequency
Vf is relative propagation velocity
maximum unaliased range is in meters
Refer to Figure 2-10, page 2-38, for a plot of maximum
range versus frequency span.
RANGE ERROR:
TOP<=BOTTOM
The SWR scale parameter top value is less
than or equal to its bottom value.
RANGE ERROR:
TOP>=BOTTOM
The RL scale parameter top value is greater
than or equal to its bottom value.
CAL
INCOMPLETE
A complete open, short, and load calibration
must be performed before calibration can be
turned on.
CAL VALID FROM:
xxxx.x to
xxxx.x MHz
The calibration was performed at a fre-
quency range that is different from the cur-
rent range.
Table 2-4.
Range Error Messages (1 of 2)
Содержание Site Master S330A
Страница 3: ......
Страница 15: ...This page is intentionally blank 1 7 Chapter 1 General Information...
Страница 23: ...2 7 Chapter 2 Operation Figure 2 3 Site Master Menu Structure...
Страница 54: ...2 38 Chapter 2 Operation Figure 2 10 Maximum Distance and Resolution vs Frequency Span...
Страница 83: ...3 14 Chapter 3 Screen Capture Program NOTES...
Страница 85: ...Color Cover P N 00986 00018...