Step 12.
Press the MKRS soft key to go to the Markers Menu
(page 2-10).
Step 13.
Press the M1 soft key.
Step 14.
Enter the desired numerical value using the keypad or
the Up/Down Arrow key. Press
ENTER
when data
entry is complete.
Step 15.
Repeat the above two steps for M2.
Step 16.
Press the MAIN soft key to return to the Main Menu.
Step 17.
Press the DIST soft key to go to the Distance Menu.
Step 18.
Press the MORE soft key to go to the Distance Sub-
Menu.
Step 19.
Press the CABLE soft key to select a coaxial cable
from the menu or press the LOSS and PROP V soft
keys to enter cable loss and relative propagation ve-
locity parameters manually (page 2-12). The loss of
the cable shown in the cable menu is @1000 MHz.
Step 20.
Enter the desired numerical values using the keypad
or the Up/Down Arrow key. Press
ENTER
when
data entry is complete.
Step 21.
Press the MAIN soft key to return to the Main Menu.
Step 22.
After calculations are complete (approximately 8 sec-
onds), observe that the message area displays the re-
turn loss data at the two marker points.