Adjusting a Limit
Step 15.
Press the
LIMIT
key (Figure 2-1, page 2-0) to call up
the Scale Menu.
Step 16.
If the Limit needs to be turned OFF, press the LIMIT
soft key on the Scale Menu. If the Limit needs to be
turned ON, press the LIMIT soft key again. This will
open the Limit parameter for entry. If needed, enter
the desired numerical value using the keypad or the
Up/Down Arrow key (Figure 2-1, page 2-0). Press
ENTER when the data entry is complete.
Step 17.
If you wish to have an audible beep when the trace is
above the limit line, press the OPT soft key from the
Main Menu (page 2-8).
Step 18.
Press the B2 soft key from the Option Menu (page
2-17) to toggle the limit beep indicator
ON
.
Saving a Setup
Step 19.
Press the
SAVE SETUP
key (Figure 2-1, page 2-0).
Step 20.
Enter the desired numerical value using the keypad (1
to 9) or the Up/Down Arrow key. Press
ENTER
when data entry is complete.
Recalling a Setup
Step 21.
Press the
RECALL SETUP
key.
Step 22.
Enter the desired numerical value using the keypad (1
to 9) or the Up/Down Arrow key. Press
ENTER
2 - 28