2-15
Power Meter Verification (Option 29)
Performance Verification
2-28
PN: 10580-00195 Revision B
S312D MM
2-15
Power Meter Verification (Option 29)
The test in this section can be used to verify the level accuracy of the Internal Power Meter (Option 29) of the
Site Master.
Required Equipment (refer to
):
•
Anritsu Model MG3692x Synthesized Signal Generator
•
Power Meter, Anritsu Model ML2438A
•
Site Master S312D
•
Power Splitter, Aeroflex/Weinschel Model 1870A
•
Power Sensor, Anritsu Model MA2442D
•
Fixed Attenuator, Aeroflex/Weinschel Model 44-10
•
Adapter, Anritsu Model 34RKNF50
•
Adapter, Anritsu Model 34NN50A
•
RF Coaxial Cable, Anritsu Model 15NN50-1.5C
Figure 2-13.
Power Meter Verification (Option 29) Characterization Setup
Sensor B
Sensor A
1870A
Power Splitter
ML2438A Power Meter
MG369XA Source
15NNF50-1.5B
34RKNF50
Adapter
Attenuator
10 dB