2-8
Spectrum Analyzer Resolution Bandwidth Accuracy Verification
Performance Verification
2-14
PN: 10580-00195 Revision B
S312D MM
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
1 kHz
. Press
ENTER
to set the
resolution bandwidth to 1 kHz and press
BACK
.
5.
Press the
Measure
soft key then the
MARKER
key.
6.
Press the
M2
soft key. Record the
Δ
2 frequency in
300 Hz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter
450
and press the
Hz
soft key to set the span to 450 Hz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
1 kHz
. Press
ENTER
to set the
resolution bandwidth to 300 Hz and press
BACK
.
5.
Press the
Measure
soft key then the
MARKER
key.
6.
Press the
M2
soft key. Record the
Δ
2 frequency in
.
100 Hz RBW Test
1.
Press the
FREQ/DIST
key.
2.
Press the
Span
soft key, enter
150
and press the
Hz
soft key to set the span to 150 Hz.
3.
Press the
MEAS/DISP
key and the
Bandwidth
soft key.
4.
Press the
RBW Manual
soft key and use the
Up/Down
arrow key to select
100 Hz
. Press
ENTER
to set the
resolution bandwidth to 100 Hz and press
BACK
.
5.
Press the
Measure
soft key then the
MARKER
key.
6.
Press the
M2
soft key. Record the
Δ
2 frequency in
Содержание Site Master S312D
Страница 2: ......
Страница 10: ...1 4 Replaceable Parts General Information 1 4 PN 10580 00195 Revision B S312D MM ...
Страница 48: ...3 7 RTC Battery Information Removal and Replacement 3 8 PN 10580 00195 Revision B S312D MM ...
Страница 58: ...Test Records A 10 PN 10580 00195 Revision B S312D MM ...
Страница 60: ...Test Fixture Schematics B 2 PN 10580 00195 Revision B S312D MM ...
Страница 61: ......