![Ankom TDF28 Скачать руководство пользователя страница 2](http://html.mh-extra.com/html/ankom/tdf28/tdf28_service-procedure_2944725002.webp)
In-Line Heater Replacement
(TDF28/28.1)
TDF
ANKOM Technology ● 2052 O’Neil Road ● Email: [email protected] ● Phone: 315-986-8090 ● Fax: 315-986-8091
Service procedure
TS018
Rev. B
6/04/2018
Page 2 of 10
6. Detach the white 3-pin connector
labeled “J10” from the MUX-T circuit
board.
5. Remove the 6 tubes going from the
downward poin<ng brass tubes on
the In-Line Heater to the Pressure
Sensor. Only remove the ends of the
tubes that are connected to the
Pressure Sensor barbs (Red Arrows).
7. Using a dental pick or a small
screwdriver, depress the tabs (Black
Arrow) inside the J10 connector in
order to slide the two conductor out
of the terminals (Red Arrows). Set
aside the white J10 connector for
later use.
8. Loosen the black nut on the explosion
proof
fiIng,
located
on
the
underside of the Electrical Enclosure,
that the white thermistor wire runs
through un<l the nut slides off.