In-Line Heater Replacement
(TDF28/28.1)
TDF
ANKOM Technology ● 2052 O’Neil Road ● Email: [email protected] ● Phone: 315-986-8090 ● Fax: 315-986-8091
Service procedure
TS018
Rev. B
6/04/2018
Page 10 of 10
37. Perform a QC test to ensure the func<onality
of the newly installed In-Line Heater.
a. Plug in the instrument and turn the power
back on
b.
CAUTION:
At this point the instrument is open
and energized. Keep personnel away from
open wires in the Electrical Enclosure.
b. On the bo=om of the “Select a Func<on”
screen, press the “Diagnos<cs” bu=on.
c. On the “Diagnos<cs” screen, press the “Heater
Test” bu=on.
d. Once in the “Heater Test” screen, check that
the In-Line Heater temperature is reading at
room temperature.
e. Press the “In-Line Heater” bu=on to turn it on.
Temperature will start to rise. Once it reaches
50°C, press again to turn off.
39. Reinstall the top stainless steel panels and the
clear plas<c back panel using the screws, nuts,
and panels set aside from earlier.
(on)
38. With cau<on, lightly tap the newly installed
In-Line Heater in the back of the instrument to
manually confirm that it is warming up. With
this completed, you are now ready to return
your ANKOM Dietary Fiber Analyzer to
service.
TDF