At least one of the valid probes selected must be an Activation probe or the User will not be
able to progress to the next screen. Once connected, the system will take up to five seconds
to determine probe validation. If the probe is valid it will have a working time of eight (8)
hours, at which time they will become invalid. Once activated a probe cannot be used on
another system even if there is still working time left.
The circle indicated in
Blue
shows that the Activation probe is connected and valid. The
circle indicated in
Green
shows that the standard probe is connected and valid, Figure 5-16.
The
Red
circle indicates that the probe is invalid or expired and the Grey circle indicates that
the probe is not connected to the Generator, Figure 5-17.
Invalid (Red)
Not Connected (Grey)
Figure 5-17: Invalid & Not Connected Probes
NOTE: The User cannot progress to the next screen without at least one valid Activation
probe connected to the Generator.
Once the desired probe configuration is determined and the corresponding number of
probes matches the configuration and an activation probe is connected, select the
Next
button
to advance to the Probe Placement Process Screen. If the User tries to advance to
the next screen without an activation probe connected, a pop-up warning window (see
Figure 5-18) will appear that reads "
Probes Selecting Error
−
No activation Probe
selected."
Connect a valid activation probe (blue handle) to proceed to the next screen.
Figure 5-18: Probe Placement Process screen Error Message
NOTE 1: More than one Activation probe may be connected.
NOTE 2: The User may connect more probes than the number of probes selected. In this
case, the User must select which probes to use by activating the check box to the left of the
probe circle.
Содержание NanoKnife
Страница 1: ......
Страница 6: ...6 160 104694 Rev 01 Version 2 2 0...
Страница 16: ...16 160 104694 Rev 01 Version 2 2 0...
Страница 26: ...26 160 104694 Rev 01 Version 2 2 0...
Страница 30: ...30 160 104694 Rev 01 Version 2 2 0...
Страница 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Страница 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Страница 70: ...70 160 104694 Rev 01 Version 2 2 0...
Страница 78: ...78 160 104694 Rev 01 Version 2 2 0...
Страница 86: ...86 160 104694 Rev 01 Version 2 2 0...