Whenever necessary, replace Generator fuses only with spare fuses specified
in this manual, see Section.
•
Maintenance should be carried out only by trained personnel. The Generator
must undergo periodic preventative maintenance as specified in the
Maintenance and Service section.
•
The NanoKnife User Manual is a fundamental part of the Generator and should
always accompany it. Users must refer to this manual for correct and complete
information on the use of the Generator.
2.5 Precautions
•
Electrodes that are not parallel to each other may result in an incomplete
ablation.
•
Inappropriately positioned electrodes or metal implants in the field may distort
the desired IRE ablation field.
•
Read this User Manual thoroughly before operating the Generator. Do not
hesitate to contact your local supplier or the manufacturer in case of doubt on
the correct usage of the Generator.
•
Electrodes are subjected to potentially harmful electric energy. Do not touch
the metal part of the electrodes while a procedure is in progress.
•
The effects of IRE on a fetus are not known. Procedure on pregnant women
should be contemplated only after ensuring that the procedure benefits
outweigh the risks.
•
Procedure safety and efficacy may be affected if electrodes other than those
supplied by AngioDynamics or by an authorized distributor are used.
•
Unless there is a reasonable doubt that a site has been ablated ineffectively,
repeating an ablation on the same site is not advisable, since it is not believed
to increase procedure efficacy.
•
Avoid unnecessarily high voltage or excessive number of pulses.
•
Use of operator-defined parameters increases the risk of ineffective procedures
or post-procedure complications, with respect to validated standard
procedures.
•
Avoid short-circuiting the electrodes when delivering pulses. Electrode to
electrode contact or electrode to electrode spacing less than 5 mm
(millimeters) may result in short circuiting during energy delivery resulting in
incomplete ablation.
•
Ensure the Generator is connected to the proper main power supply value (see
Section 9.4 and that the main power supply outlet is able to supply the required
power.
•
Do not use the Generator if a malfunction is suspected. Contact the
manufacturer or the local authorized supplier.
•
Avoid intentional or accidental spilling of liquids on the Generator. In particular,
do not keep containers of liquids on the Generator. Do not handle the
equipment with wet hands.
Содержание NanoKnife
Страница 1: ......
Страница 6: ...6 160 104694 Rev 01 Version 2 2 0...
Страница 16: ...16 160 104694 Rev 01 Version 2 2 0...
Страница 26: ...26 160 104694 Rev 01 Version 2 2 0...
Страница 30: ...30 160 104694 Rev 01 Version 2 2 0...
Страница 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Страница 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Страница 70: ...70 160 104694 Rev 01 Version 2 2 0...
Страница 78: ...78 160 104694 Rev 01 Version 2 2 0...
Страница 86: ...86 160 104694 Rev 01 Version 2 2 0...