4.1 System
Location
The Generator must be installed and operated in an
environment that conforms to the operating conditions specified
in Section 10.4.
The Generator must be installed on rigid surfaces suitable to
withstand its weight.
In addition, the Generator must be installed in a way that any
surface parallel to the Power Unit’s rear panel and in
relationship to its ventilation grids, remain at least 5 cm
(centimeters) away.
Care must be taken to avoid items (e.g., dust covers) that can
occlude the ventilation grids.
WARNING!
The Generator must be connected to the safety ground of
the main. The
Main Power Supply Cord
can ensure proper
grounding only if connected to properly grounded main
system, through suitable main sockets.
WARNING!
The Generator must not be used in the presence of
flammable mixtures.
Installation Instructions
•
Connect the
Main Power Supply Cord
(supplied by
the manufacturer) to the cord connector located on
the back panel.
S E C T I O N
Содержание NanoKnife
Страница 1: ......
Страница 6: ...6 160 104694 Rev 01 Version 2 2 0...
Страница 16: ...16 160 104694 Rev 01 Version 2 2 0...
Страница 26: ...26 160 104694 Rev 01 Version 2 2 0...
Страница 30: ...30 160 104694 Rev 01 Version 2 2 0...
Страница 42: ...42 160 104694 Rev 01 Version 2 2 0 Three Probe Array Four Probe Array Five Probe Array...
Страница 43: ...43 160 104694 Rev 01 Version 2 2 0 Six Probe Array Six Probe Array 10 mm Six Probe Array 15 mm...
Страница 70: ...70 160 104694 Rev 01 Version 2 2 0...
Страница 78: ...78 160 104694 Rev 01 Version 2 2 0...
Страница 86: ...86 160 104694 Rev 01 Version 2 2 0...