Dealing with “Non-Leak Connections to Ground”: 

 

Although it is always best to remove connections to ground that are not leaks  

either physically (as in removing ladders) or by covering (as with light cover), this 

is not always possible. There will be situations where the strongest “Leak Track” 

in the pool is actually going to a light or stair gasket screw that is not necessarily 

leaking. To find leaks in these situations move the probe close to the non-leak 

connection to ground, then turn the striped end of the probe away and survey the 

vinyl moving away from the non-leak connection. If you pick up a signal pointing 

away from the strong non-leak connection it is probably a leak. 

 

In situations where there are other connections to ground such as mentioned 

above it will be necessary to survey the entire pool in a tighter pattern than would 

be necessary is such other connections were not present. 

 

When checking for leaks in close proximity to “non-leak connections” use the 

“low” setting. 

 

 

Содержание LeakTrac Wireless 2400

Страница 1: ...INSTRUCTION MANUAL 2885 Country Dr 190 St Paul MN 55117 800 348 1316 www leaktools com Your Partner in Swimming Pool Leak Detection 2400...

Страница 2: ...no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be deter mined by turning the eq...

Страница 3: ...ck The sophisticated LeakTrac 2400 detects these Leak Tracks by measuring the differential voltage potential at different points in the pool and translates this in formation into audible clicking soun...

Страница 4: ...s neck Touch pad buttons and LED indicators offer the following user interface SENSITIVITY Use the up arrow button to toggle through three different input am plification settings Current setting will...

Страница 5: ...1 Red accepts the banana plug connector from the red Ground Cable 2 Black accepts the banana plug connector from the Input Float Cable A toggle switch turns the Booster on I and off 0 A LED indicator...

Страница 6: ...o make a connection to ground A length of insulated wire to red banana plug is used to connect to the red input jack on the Power Booster Headphones These standard headphones plug into the back of the...

Страница 7: ...to the pool the strongest Leak Track will be produced when there is only one path for the charge to connect to ground 3 Relative conductivity of different points of low resistance One weak Leak Track...

Страница 8: ...c 2400 Preliminary Steps Before using the LeakTrac 2400 first confirm that there is indeed a leak in the pool by using the Leakalyzer or doing a bucket test Next pressure test the plumbing lines and r...

Страница 9: ...over can be removed by placing the hose without the pump into the water below the bottom edge of the cover Warning This light cover is not intended for any use where the rubber inner tube gasket conta...

Страница 10: ...tach the clamp to ground Good connections to ground include Ladder anchors sockets Diving board anchor bolts Metal junction box conduit Any metal rod stuck into the ground If the ground you choose is...

Страница 11: ...electrical leakage is occurring If no variation in click rate occurs move the probe a few feet and repeat Note The Booster will turn off automatically one minute after the SPU has been turned off The...

Страница 12: ...ons Generally the signal will be most sharp at the highest possible power and click rate setting that gives rapid clicking when pointed toward the leak but no signal when pointed away User preferences...

Страница 13: ...is not necessarily leaking To find leaks in these situations move the probe close to the non leak connection to ground then turn the striped end of the probe away and survey the vinyl moving away from...

Страница 14: ...ly inserted into the jack on the back of the SPU Assure that batteries are good in Booster Check probe wire for wear on insulation Call Anderson Manufacturing Possible Cause Booster not working proper...

Страница 15: ...too much signal due to other ground connection Solution Turn Power down and re adjust Click Rate Cover lights remove ladders etc Assure liner above water line is not wet creating connection to coping...

Страница 16: ...y clean the plate with steel wool to keep it conductive Care As with any electrical equipment moisture will cause corrosion problems Do what you can to keep the equipment and the inside of the case dr...

Отзывы: