Table 1 

EMISSION Limits

 

Phenomenon Compliance 

Conducted and radiated RF EMISSION

   

 

 

CISPR 11 

Group 1

Class B 

Table 2 

IMMUNITY TEST LEVELS

Enclosure Port

 

Phenomenon 

IMMUNITY TEST LEVELS 

Electrostatic discharge 
IEC 61000-4-2 

±8 kV contact   
±2 kV, ±4 kV, ±8 kV, ±15 kV air 

Radiated RF EM fields 
IEC 61000-4-3 

10 V/m 
80 MHz - 2.7 GHz 
80 % AM at 1 kHz 

Proximity fields from RF wireless communications equipment 
IEC 61000-4-3 

See table 3 

Rated power frequency magnetic fields 
IEC 61000-4-8 

30 A/m 
50 Hz or 60 Hz 

Table 3 

Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications 

equipment

 

Test 

frequency 

(MHz) 

Band 

(MHz) 

Service Modulation 

Maximum 

power 

(W) 

Distance 

(m) 

IMMUNITY 

TEST 

LEVEL 

(V/m) 

385 

380 -  390 

TETRA 400 

Pulse modulation 

18 Hz 

1.8 0.3  27 

450 

430 - 470 

GMRS 460 

FRS 460 

FM 

±5 kHz deviation 

1 kHz sine 

2 0.3 28 

710 

704 - 787 

LTE Band 13

17 

Pulse modulation 

217 Hz 

0.2 0.3  9 

745 
780 

810 

800 - 960 

GSM 800/900 

TETRA 800 

iDEN 820 

CDMA 850 

LTE Band 5 

Pulse modulation 

18 Hz 

2 0.3 28 

870 

930 

1720 

1700 -  1990 

GSM 1800 

CDMA 1900 

GSM 1900 

DECT 

LTE Band 1

3

4

25 

UMTS 

Pulse modulation 

217 Hz 

2 0.3 28 

1845 

1970 

2450 2400 

2570 

Blutooth 

WLAN 802.11 b/g/n 

RFID 2450 

LTE Band 7 

Pulse modulation 

217 Hz 

2 0.3 28 

5240 

5100 -  5800 

WLAN 802.11 a/n 

Pulse modulation 

217 Hz 

0.2 0.3  9 

5500 
5785 

 

EMD Technical Data Battery-operated Blood Pressure Monitor 

Содержание UT-201BLE

Страница 1: ...01BLE A Oral type Instruction manual Original Type oral Manuel d instructions Traduction Tipo oral Manual de instrucciones Traducci n Tipo orale Manuale di istruzioni Traduzione Oraler typ Bedienungsa...

Страница 2: ...Units 11 Applying the Thermometer 12 Measurement Time 12 After Measurement 12 Measurements 13 Adjusted Mode Measurement 13 Direct Mode Measurement 14 Notes for Accurate Measurement 14 Features 15 Adj...

Страница 3: ...mark of conformity 0123 The reference number to the involved notified body This device fulfills the provisions of BS EN 12470 Clinical thermometers Part 3 Performance of compact electrical thermomete...

Страница 4: ...not water resistant Prevent rain sweat and water from soiling the device Measurements may be distorted if the device is used close to televisions microwave ovens X ray or other devices with strong ele...

Страница 5: ...generation or bursting and resulting in injury Do not heat the battery Failure to do so may lead to fluid leakage bursting and resulting in injury Prevent the device from chewing and bending It may b...

Страница 6: ...mmunication mark Memory mark Battery indicator Count down indicator for measurements Adjusted Mode measurement mark Direct Mode measurement mark ON Standby switch with LED Battery CR2032 Case This man...

Страница 7: ...unction of thermometer Contact your dealer Time out of Bluetooth communication Bluetooth communication error Bluetooth communication mark Pair mark to construct Bluetooth communication pair Adjusted M...

Страница 8: ...aution Insert the battery as shown in the battery compartment If installed incorrectly the device will not work When Low battery mark blinks in the display replace the battery with a new one Replace t...

Страница 9: ...aves in locations where there are many obstructions or in other locations where signal strength is weak Doing so may result in frequent loss of connection very slow communication speeds and errors Usi...

Страница 10: ...ooth devices Continua certified devices iPhone iPad iPod iPhone 4S or later Applications and devices that are compatible with Bluetooth 4 0 Each device needs an application to receive data For connect...

Страница 11: ...ceiver device to switch it to the pairable status When pairing this thermometer place it as close as possible to the receiver device to be paired with 2 Install the battery as described on page 7 Pres...

Страница 12: ...deleted and the new data is stored The amount of data that can be stored temporarily may vary with the application Time This thermometer has a built in clock The date and time that a measurement was t...

Страница 13: ...e sensor at a correct position in the mouth When the direct mode measurement is used keep the sensor position for approximately 5 minutes with placing the sensor at a correct position in the mouth We...

Страница 14: ...t down indicator is displayed The adjusted mode measurement mark blinks and rotates 6 The result is displayed for approximately 15 seconds the adjusted mode measurement mark lights the LED of the swit...

Страница 15: ...f function with the buzzer that the device is turned off at approximately one minute later from removing it or displaying data The device can be turned off when pressing and holding the switch In meas...

Страница 16: ...ishing the measurement this switch flashes Last Reading Display and Memory The previous reading stored in memory is automatically displayed when turning on the thermometer The new reading is stored in...

Страница 17: ...ll change at awaking in activity after eating Measure the temperature under the same condition Body temperature is displayed low Incorrect sensor position Check the sensor position Body temperature is...

Страница 18: ...isinfectant ethanol 76 9 to 81 4 v v Storage Store the device with avoiding extremes in temperature humidity direct sunlight vibration shock dust or fire Keep it put into the case with dry air and roo...

Страница 19: ...justed mode measurement Approx 350 times Direct mode measurement Approx 120 times Useful life 5 years Wireless communication VZ MURATA Manufacturing Co Ltd Bluetooth Ver 4 0 low energy HTP Frequency b...

Страница 20: ...RF wireless communications equipment IEC 61000 4 3 See table 3 Rated power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz Table 3 Test specifications for ENCLOSURE PORT IMMUNITY to RF w...

Страница 21: ...English 20 MEMO...

Страница 22: ...se California 95131 U S A Telephone 1 408 263 5333 Fax 1 408 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D...

Отзывы: