English 15
Features
Adjusted Mode Measurement
The adjusted mode
measurement calculates the
direct mode temperature
after five minutes when the
body temperature is
measured for 30 seconds. It
is based on an ascending
thermal curve.
If you need a precision data,
we recommend to use the
direct mode measurement for
a precision thermometry.
Direct Mode Measurement
The
body
temperature
can be measured using for five minutes.
Switch with Flash Action
When pressing the switch , this switch flashes.
When finishing the measurement, this switch flashes.
Last Reading Display and Memory
The previous reading stored in memory is automatically displayed when
turning on the thermometer. The new reading is stored in memory when
measuring the temperature.
Temperature Unit
Temperature unit of Celsius or Fahrenheit can be selected .
Bluetooth
Temperature data can be transmitted to the receiver that is paired with the
thermometer.
Results of Clinical Evaluation
Clinical Bias
-0.06 °C
Clinical Repeatability
0.12 °C
Limit of Agreement
0.22 °C
Standard Deviation
0.08 °C
Direct mode reading
Body temperature
Room temperature
30 seconds
5 minutes
Adjusted mode
reading
Thermal curve
in ascending.
Adjusted mode
calculation