English 11

 

Using the Thermometer

 

 

Measurement and Transmitting Data 

 

The communication performs the following steps after building the paring.   
Keep the condition of the receiver device so as to communicate. 

1.  Turn on the thermometer. Data is measured automatically. 
2.  Data is transmitted after finishing the measurement. 
 

Transmitting Temporarily Stored Data 

 

In cases when the receiver device cannot receive measurement data, the 
measurement data is temporarily stored in the thermometer memory. The 
data stored in the memory is transmitted the next time a connection is 
successfully made to the receiver device 
A total of 90 sets of measurement data can be stored. When the amount of 
data exceeds 90 sets, the oldest data is deleted and the new data is stored. 

 

The amount of data that can be stored temporarily may vary with the 
application.  

 

Time 

 

This thermometer has a built-in clock. The date and time that a 
measurement was taken is included in the measurement data. 

 

The built-in clock is designed to be automatically adjusted by syncing with   
the clock of a receiver device. Refer to the specifications of the receiver 
device. This thermometer has no clock adjustment function. 

 

Changing Units 

1.  Press and hold the 

 

 

  switch above 6 seconds when turning off the 

thermometer. The new unit is displayed after blinking the last unit.   
The thermometer turns off automatically. 

2.  When the same operation is 

performed again, an effective unit 
is exchanged. 

 
 

The unit is stored in the memory. 
The unit of the factory setting 
is °C (centigrade). 

 
 

 

Turn off the device. 

Press and hold the       

  switch above six seconds. 

Last unit 

 

 

 

New unit 

The device is turned off automatically. 

Содержание UT-201BLE

Страница 1: ...01BLE A Oral type Instruction manual Original Type oral Manuel d instructions Traduction Tipo oral Manual de instrucciones Traducci n Tipo orale Manuale di istruzioni Traduzione Oraler typ Bedienungsa...

Страница 2: ...Units 11 Applying the Thermometer 12 Measurement Time 12 After Measurement 12 Measurements 13 Adjusted Mode Measurement 13 Direct Mode Measurement 14 Notes for Accurate Measurement 14 Features 15 Adj...

Страница 3: ...mark of conformity 0123 The reference number to the involved notified body This device fulfills the provisions of BS EN 12470 Clinical thermometers Part 3 Performance of compact electrical thermomete...

Страница 4: ...not water resistant Prevent rain sweat and water from soiling the device Measurements may be distorted if the device is used close to televisions microwave ovens X ray or other devices with strong ele...

Страница 5: ...generation or bursting and resulting in injury Do not heat the battery Failure to do so may lead to fluid leakage bursting and resulting in injury Prevent the device from chewing and bending It may b...

Страница 6: ...mmunication mark Memory mark Battery indicator Count down indicator for measurements Adjusted Mode measurement mark Direct Mode measurement mark ON Standby switch with LED Battery CR2032 Case This man...

Страница 7: ...unction of thermometer Contact your dealer Time out of Bluetooth communication Bluetooth communication error Bluetooth communication mark Pair mark to construct Bluetooth communication pair Adjusted M...

Страница 8: ...aution Insert the battery as shown in the battery compartment If installed incorrectly the device will not work When Low battery mark blinks in the display replace the battery with a new one Replace t...

Страница 9: ...aves in locations where there are many obstructions or in other locations where signal strength is weak Doing so may result in frequent loss of connection very slow communication speeds and errors Usi...

Страница 10: ...ooth devices Continua certified devices iPhone iPad iPod iPhone 4S or later Applications and devices that are compatible with Bluetooth 4 0 Each device needs an application to receive data For connect...

Страница 11: ...ceiver device to switch it to the pairable status When pairing this thermometer place it as close as possible to the receiver device to be paired with 2 Install the battery as described on page 7 Pres...

Страница 12: ...deleted and the new data is stored The amount of data that can be stored temporarily may vary with the application Time This thermometer has a built in clock The date and time that a measurement was t...

Страница 13: ...e sensor at a correct position in the mouth When the direct mode measurement is used keep the sensor position for approximately 5 minutes with placing the sensor at a correct position in the mouth We...

Страница 14: ...t down indicator is displayed The adjusted mode measurement mark blinks and rotates 6 The result is displayed for approximately 15 seconds the adjusted mode measurement mark lights the LED of the swit...

Страница 15: ...f function with the buzzer that the device is turned off at approximately one minute later from removing it or displaying data The device can be turned off when pressing and holding the switch In meas...

Страница 16: ...ishing the measurement this switch flashes Last Reading Display and Memory The previous reading stored in memory is automatically displayed when turning on the thermometer The new reading is stored in...

Страница 17: ...ll change at awaking in activity after eating Measure the temperature under the same condition Body temperature is displayed low Incorrect sensor position Check the sensor position Body temperature is...

Страница 18: ...isinfectant ethanol 76 9 to 81 4 v v Storage Store the device with avoiding extremes in temperature humidity direct sunlight vibration shock dust or fire Keep it put into the case with dry air and roo...

Страница 19: ...justed mode measurement Approx 350 times Direct mode measurement Approx 120 times Useful life 5 years Wireless communication VZ MURATA Manufacturing Co Ltd Bluetooth Ver 4 0 low energy HTP Frequency b...

Страница 20: ...RF wireless communications equipment IEC 61000 4 3 See table 3 Rated power frequency magnetic fields IEC 61000 4 8 30 A m 50 Hz or 60 Hz Table 3 Test specifications for ENCLOSURE PORT IMMUNITY to RF w...

Страница 21: ...English 20 MEMO...

Страница 22: ...se California 95131 U S A Telephone 1 408 263 5333 Fax 1 408 263 0119 A D AUSTRALASIA PTY LTD 32 Dew Street Thebarton South Australia 5031 AUSTRALIA Telephone 61 8 8301 8100 Fax 61 8 8352 7409 OOO A D...

Отзывы: