background image

ADM, HBOT & Sub Aspida 
User Manual 

Int. Approved   

 

Document ref: PSA-811-14 

June 2017 

Page 28 of 43 

 

Copyright © 2017 Analox Ltd. All Rights Reserved. 

 

Commercial in Confidence 

6.4

 

Oxygen sensor replacement 

The  oxygen  sensor  used  in  oxygen  versions  of  the 

Aspida

  is  an  electrochemical  cell,  and  hence  the 

sensor  output  will  deplete  over  time,  the  expected  life  of  the  sensor  is  1  year  (  With  a  6  month 
warranty), depending on usage.  After a 1 year period of service, or when the sensor has depleted, the 

oxygen sensor should be replaced with a fresh sensor. (Contact supplier for a replacement sensor). 

 

 

WARNING:

 Before  replacing the sensor,  ensure the device is switched off (disconnect from 

any USB cable or charger). Open the battery compartment and remove the batteries from the 

device. 

WARNING

The  sensor  in  the 

Aspida

  is  an  electrochemical  device  and  contains  a  caustic 

electrolyte.  The  sensors  are  themselves  sealed  and  do  not  under  normal  circumstances 

present  a  health  hazard  however  if  leakage  of  the  Potassium  Hydroxide  electrolyte  has 

occurred use rubber gloves and wear chemical splash goggles to handle and clean up and do 
not allow the electrolyte onto any part of your body or clothing. Rinse contaminated surfaces 

with  water.    In  the  event  that  you  do  come  into  contact  with  the  electrolyte  wash  the 

contaminated part with copious amounts of water 

 
For ADM Aspida

’s sec

tion 4.2 should be followed to gain access to the instrument. 

 

To replace the oxygen sensor: 
 

1) 

Remove the top 4 screws (see Figure 2) from the device using a cross-head screwdriver as 

shown in Figure 16.  Retain the screws along with the sealing washers. 

 

Figure 16- Remove the sensor cap screws 

 

Содержание Sub Aspida

Страница 1: ...Business Park North Yorkshire TS9 5PT UK UK RoW T 44 0 1642 711400 F 44 0 1642 713900 US T 714 891 4478 W www analox net E info analox net Copyright 2017 Analox Ltd All Rights Reserved ADM HBOT Sub A...

Страница 2: ......

Страница 3: ...registration 14 5 4 The main display 15 5 5 Screen saver 15 5 6 Battery status 16 5 7 Menus 16 5 8 Common menu items 17 5 9 Gas alarms 18 5 10 Global alarm options 19 5 11 Alarm latching 19 5 12 Alarm...

Страница 4: ......

Страница 5: ...nit should always be worn away from the head in order to minimise close range exposure to the alarm WARNING The 110dB man down alarm can when enabled be manually activated at any time by holding the b...

Страница 6: ...teries c Wall mounting plate Including fixing kit or belt clip attachment only one supplied d Charger power supply to suit e USB communication cable f Software disc g Quick start guide h Calibration a...

Страница 7: ...June 2017 Page 5 of 43 Copyright 2017 Analox Ltd All Rights Reserved Commercial in Confidence ADM Aspida a ADM panel mount main unit b Universal charger power supply Including adaptors c Quick start...

Страница 8: ...than 12 hours continuously between charges An Analox Aspida allows for easy replacement of the rechargeable batteries with standard AA type non rechargeable batteries in circumstances where rechargin...

Страница 9: ...mbar ppO2 versions 18 19 5 23 180 mbar ppO2 195 mbar ppO2 230 mbar ppO2 CO2 Alarms SEV mbar ppCO2 versions 0 5 1 5 4 5 mbar ppCO2 15 mbar ppCO2 40 mbar ppCO2 5 mbar ppCO2 TWA Alarm Alarm Latch State...

Страница 10: ...features Figure 1 Main features 1 Alarm fault and OK indicators 2 Carbon dioxide gas port 3 Cancel exit panic alarm button 4 Cycle button 5 Confirm on off button 6 Oxygen gas port if sensor fitted 7 O...

Страница 11: ...will be complete within approximately 4 5 hours 4 1 Installation of the ADM Aspida The ADM Aspida will come with the battery cover secured to the main unit and in turn secured to the front panel The...

Страница 12: ...align the ADM Aspida front panel mounting holes up with the captive nuts then re insert the M5 x 25mm Pozi Pan screws through and tighten The diagram below shows the exploded view of the ADM installat...

Страница 13: ...positive Battery backup is available as an optional extra this can be supplied by Analox see spares and accessories list at section 6 1 To gain access to the battery compartment undo the screw in the...

Страница 14: ...da may be used with standard rechargeable AA type batteries which have been charged using a third party charging device However only NiMH batteries provided by Analox should be used when attempting to...

Страница 15: ...d None Charging charge setup animated None Charging animated None Charging approx 1 5 hours remain Success beep Charge complete flashing Fault beep Charge fault Batteries not rechargeable Fault beep N...

Страница 16: ...lay press and hold the button until the screen goes blank then release the button To return to the main display from the menu system press and release the button as required to exit the menus NOTE The...

Страница 17: ...asured gas units single sensor NOTE The ADM Aspida does not include a pressure sensor it has a fixed atmospheric pressure reading of 1000 mbar This value is displayed as above but will remain at 1000...

Страница 18: ...TE Due to the nature of the battery type the battery status icon may temporarily indicate a capacity higher than the true remaining battery capacity upon switch on This effect is usually seen in devic...

Страница 19: ...only Select this option to perform a calibration of the carbon dioxide sensor This feature is described in further detail in the section 6 3 Oxygen calibration oxygen configurations only Select this o...

Страница 20: ...g that carbon dioxide levels have triggered an alarm Each alarm is configurable as either high going or low going In the case of a high going alarm if the gas value exceeds the configurable set point...

Страница 21: ...e set point In the case of multiple alarms the highest priority alarm that was triggered will be latched If the priority 1 alarm is latched and the gas level drops below the priority 1 set point press...

Страница 22: ...Commercial in Confidence 5 13 Quiet alarms Sometimes it is desirable to suppress the audible indication when alarm is triggered The Analox Aspida provides an option to allow alarms to be silenced The...

Страница 23: ...If the man down feature is enabled the device will sound a low volume pre warning alarm after it detects no movement over a period of 4 minutes The pre warning alarm will sound for a period of 1 minu...

Страница 24: ...When a user powers on the device and selects a user ID the Analox Aspida begins calculating and continuously updating that user s own TWA The device also takes into account any historic data from the...

Страница 25: ...flashing below the reading of the sensor that is due calibration as shown in Figure 9 This symbol will continue to flash until a calibration has been performed Figure 9 Carbon dioxide calibration due...

Страница 26: ...batteries or incorrectly orientated batteries Check that batteries are inserted correctly Device does not switch on Batteries are flat Check that the batteries are inserted correctly and are recharge...

Страница 27: ...ly 2822 0008A 9V DC US Charger For Sub HBOT Aspida s only 2822 0011A 9V DC Aust Charger For Sub HBOT Aspida s only 2822 0035 9V Universal Charger PSU For ADM Aspida only PGA 607 9V DC Car Charger P013...

Страница 28: ...ch Aspida type are shown in the table below In order for a calibration due notification to be cleared for a particular sensor the user must perform all of the calibrations available for the sensor For...

Страница 29: ...ible alert will be given and a calibration confirmation screen will be shown as below Figure 14 Calibration failed Figure 15 Calibration passed 7 If the calibration passed then the tick icon will be s...

Страница 30: ...and remove the batteries from the device WARNING The sensor in the Aspida is an electrochemical device and contains a caustic electrolyte The sensors are themselves sealed and do not under normal circ...

Страница 31: ...in Confidence 2 Gently remove the sensor cap as shown in Figure 17 Figure 17 Remove the sensor cap WARNING Do not touch the inside of the instrument other than as noted in these instructions without...

Страница 32: ...ver the oxygen sensor right hand sensor being careful not to cause damage to the carbon dioxide sensor left hand sensor if fitted Ensure that the alignment rib on the extraction tool lines up with the...

Страница 33: ...ift the extraction tool vertically away from the device and the oxygen sensor should lift out of its socket as shown in Figure 22 Figure 22 Removing the oxygen sensor 8 Remove the old sensor from the...

Страница 34: ...led bags Before the bag is opened check that the sensor has not leaked if it has then please refer to the safety information at the start of this section Figure 23 Insert the new sensor into the extra...

Страница 35: ...igure 25 Align the new sensor with the Aspida device 12 Push the tool and sensor gently downwards into the case until the sensor locates firmly within its socket 13 Hold the top ring of the extraction...

Страница 36: ...e 28 Replace the sensor cap and secure the screws WARNING Any new sensor fitted to a device will take time to settle to a stable reading For this reason once the sensor has been fitted the device shou...

Страница 37: ...eaving the unit to dry to ensure the pores of the membranes over the sensors are not blocked by dirt or salt The unit should be turned off after accidental immersion and not turned on again until it h...

Страница 38: ...e batteries Battery discharge time 12 hours under normal operation passive atmospheric monitoring minimal user interaction no alarms Battery lifespan 2 years Battery charge time 4 5 hours from flat Ch...

Страница 39: ...Accuracy at standard temperature and pressure 1 of full scale 2 of reading Oxygen sensor where fitted Sensor type Electrochemical Range 0 1 to 2000 0 mbar ppO2 0 1 to 100 0 O2 at 1000mbara Response t...

Страница 40: ...ay elect instead to provide to you which ever is the lesser of the cost of replacement or a refund of net purchase price paid as per our Invoice on initial purchase from us We shall have no liability...

Страница 41: ...declared that the following product Product names ADM Aspida HBOT Aspida Sub Aspida Product codes ADM HBOT PSA Conforms to all applicable requirements of BS EN 61000 6 1 BS EN 61000 6 3 IEC EN 61010 1...

Страница 42: ...electronic product can not be placed in household waste bins Please check local regulations for information on the disposal of electronic products in your area 10 2 Oxygen sensor disposal The oxygen s...

Страница 43: ......

Отзывы: