6.4 

Proof test 

 
To  verify  that  the  system  is  still  operating  within  specification,  you  will  need  the  following 
equipment: 
 

1)  A test gas cylinder containing 17% oxygen in nitrogen, +/-2% tolerance 
2)  A bump test gas (For Safe-

Ox+™ only) with 23.5% oxygen in nitrogen, +/-2% tolerance 

3)  0.5 lpm regulator to suit calibration gas bottle 
4)  Flow adaptor - Part No: 8000-0069A 

 

Then follow the procedure for the particular instrument type: 

6.4.1 

Ax 

O2NE+™ 

1)  Fit  the  flow  adaptor  to  the  O2  sensor  and  connect,  with  tubing,  to  the  17%  oxygen 

cylinder and open the regulator to allow for the flow of gas over the sensor. 

2)  After a few moments, the 'Alarm 1' alarm should operate, followed a little later by the 

'Alarm 2' alarm.   

3)  Remove the test gas. It will take a little while for the gas to diffuse out of the instrument 

sensor, allowing the unit to recover to its normal non-alarm condition.  

6.4.2 

Ax Safe-

Ox+™ 

1)  Fit  the  flow  adaptor  to  the  O2  sensor  and  connect,  with  tubing,  to  the  17%  oxygen 

cylinder and open the regulator to allow for the flow of gas over the sensor. 

2)  After a few moments, the 'Alarm 2' alarm should operate. 
3)  Remove the test gas. It will take a little while for the gas to diffuse out of the Instrument 

Sensor, allowing the unit to recover to its normal non-alarm condition. 

4)  Fit the flow adaptor to the O2 sensor and connect, with tubing,  to the 23.5% oxygen 

bump test cylinder and open the regulator to allow for the flow of gas over the sensor. 

5)  After a few moments, the 'Alarm 1' alarm should operate. 
6)  Remove the test gas. It will take a little while for the gas to diffuse out of the instrument 

sensor, allowing the unit to recover to its normal non-alarm condition. 

6.5 

Adjusting alarm set points 

The procedure is very similar for setting either Alarm 1 or Alarm 2. 
 

1)  Whilst the 

O2NE+™/Safe-Ox+™ is switched on, enter technician mode by pressing the 

mode switch 3 times.  If entered successfully the green LED will flash off for 1.5 seconds 
and on for 0.5 of a second. 

2) 

From technician mode, press the ‘Mode’ switch 2 times to set Alarm 1 or 3 times to set 
Alarm  2.    The  buzzer  will  bleep  on  each  press.    If  this  is  done  successfully,  the 
instrument will show the ‘Fault’ indicator and the appropriate ‘Alarm’ indicator.  If this is 
done  inadvertently,  or  if  another  mode  is  selected,  press  the  ‘Mode’  switch  once  to 
return to technician mode and then repeat this selection. 

3)  The display will indicate the present value of the alarm. 
4) 

Press the ‘Mode’ switch twice to proceed to define a new setting, or once to abort and 
return to technician mode. 

5) 

Press and hold the ‘Mode’ switch. The displayed value will count down at approximately 
one count per second.  Release the switch when the displayed value is equal to the 
desired alarm value. 

6) 

Upon release of the ‘Mode’ switch, the display  will continue to show  the new value.  
Accept the new setting by pressing the ‘Mode’ switch twice, or alternatively ignore the 
new setting by pressing the switch once.  This will return to the technician mode. 

7)  To exit from technician mode, press the Mode switch once.  The 

O2NE+™/Safe-Ox+™ 

then restarts by performing the normal power on sequence (4 flashes). 
 

Содержание O2NE+

Страница 1: ...usiness Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net O2NE Safe Ox User Manual THIS IS A PRECISION DEVICE KEEP UP RIGHT ALLOW 2 HOUR WARM UP...

Страница 2: ......

Страница 3: ...splay 6 4 2 Wall mounting 6 4 3 Wiring installation 7 4 4 AC supply models 7 4 5 DC supply models 7 4 6 Alarm repeaters 8 4 7 Alarm relay output models 10 4 8 Relay wiring 10 4 9 4 to 20mA output mode...

Страница 4: ...early to 50 relative humidity at 40 C 5 Mains voltage supply fluctuations not to exceed 10 of the nominal voltage 6 Impulse withstand over voltage category II of IEC 60364 4 443 7 Pollution degree 2 8...

Страница 5: ...nally connected equipment introducing hazardous voltages into the A50 O2NE Safe Ox or CO Clear It is the responsibility of users of this product to ensure that their equipment accessory is safe before...

Страница 6: ...On receipt of the Ax O2NE Safe Ox please check you have the following 1 Ax O2NE Safe Ox main unit and alarm repeater 2 User manual 3 Test certificate 4 Rawl plugs and screws for wall mounting 5 Drill...

Страница 7: ...s this can affect the performance of the pressure compensation sensor over time Contact Analox for installation advice if you intend to use the unit in this way 3 2 Setup Different versions of the ins...

Страница 8: ...be removed from the instrument display window This is an easy peel label and should be removed easily without leaving behind any residue 4 2 Wall mounting The Ax O2NE Safe Ox should be mounted onto a...

Страница 9: ...Safe Ox s are pre wired with a mains cable fitted with a plug suited to the destination country Where internal plug fuses are fitted these are 3 amp Ensure that the unit is connected to the correct su...

Страница 10: ...Ax O2NE Safe Ox pass through one of the cable glands on the Remote Alarm Repeater 3 Terminate the wires in accordance with the table below 1 Replace the lid of the Remote Alarm Repeater insert and ti...

Страница 11: ...ote Alarm Repeater 3 Terminate the wires in accordance with the table below 4 Connect the free wire Brown or Orange to A50 231 main PCB at JP13 terminal 5 3 Replace the lid of the Remote Alarm Repeate...

Страница 12: ...should be connected 4 8 Relay wiring The cable gland is for cables of outside diameter between 5 and 7mm if cable fitted is outside that range a suitably specified cable gland must be used Ensure tha...

Страница 13: ...Document ref P0120 820 14 August 2017 Page 11 of 23 Commercial in Confidence The customer system should be connected to this cable as follows and as per the following drawing 1 Blue wire 4 to 20mA ne...

Страница 14: ...of alarms They may be factory configured in the opposite sense if required 5 2 Alarm indications 5 2 1 Ax Safe Ox If the Ax Safe Ox detects an O2 concentration which is greater than the high alarm lev...

Страница 15: ...wn the instrument and remove from its installed location to allow access to the lid screws Remove the lid of the instrument to gain access to the internals of the instrument Disconnect the sensor from...

Страница 16: ...ses used on the Ax O2NE Safe Ox must be manufactured to a 2 tolerance Note For best performance the monitor should be powered and allowed to warm up for approximately 2 hours before calibration is per...

Страница 17: ...will illuminate sound 4 times before returning to normal operation If during the one minute wait period the red or amber LED s are illuminated consult the table below ALARM1 LAMP RED ALARM2 LAMP RED F...

Страница 18: ...w for the flow of gas over the sensor 5 After a few moments the Alarm 1 alarm should operate 6 Remove the test gas It will take a little while for the gas to diffuse out of the instrument sensor allow...

Страница 19: ...ND FAST BUZZER O2 Cell Fault Output too High OFF FLASHING FLASHING FLASHING AND FAST BUZZER O2 Cell Fault Output too Low OFF OFF OFF FLASHING AND VERY FAST BUZZER Internal system fault contact Analox...

Страница 20: ...O2NE 600g Alarm Repeater 150g Dimensions O2NE Safe Ox 175x105x75 mm Alarm Repeater 155x72x45 mm IP rating O2NE Safe Ox IP65 Alarm Repeater Hard wired IP65 Quick connect IP43 Sensor type Electro Chemi...

Страница 21: ...14 August 2017 Page 19 of 23 Commercial in Confidence 8 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Страница 22: ...O2NE Safe Ox User Manual Int Approved Document ref P0120 820 14 August 2017 Page 20 of 23 Commercial in Confidence 9 Declaration of conformity...

Страница 23: ......

Отзывы: