O2NE+ & Safe-Ox+
User Manual
Int. Approved
Document ref: P0120-820-14
August 2017
Page 13 of 23
Commercial in Confidence
6
Maintenance
6.1
Oxygen cell replacement
The following instructions describes the replacement of th
e O2NE+™ & Safe-Ox+™ oxygen
sensor.
•
Upon receiving the new sensor, remove from its packaging and leave for 24 hours to
stabilise in air.
•
Power down the instrument and remove from its installed location to allow access to
the lid screws.
•
Remove the lid of the instrument to gain access to the internals of the instrument.
•
Disconnect the sensor from its location on the PCB.
•
Remove the locking ring from the sensor and remove the sensor from the enclosure.
•
Refit the new sensor by reversing the above process.
•
Refit the enclosure lid and re-install the instrument and power up.
Upon renewing the oxygen sensor a full calibration will be required to be performed, this
should be done 2 hours after power up.