6.6 

Fault conditions 

During  normal  operation,  the  instrument  carries  out  a  continuous  self-test  procedure.    If 
operation  is  satisfactory,  the  ‘Good/OK’  status  indicator  will  be  on,  blinking  off  momentarily 
every few seconds. 
 

1)  If  there  are  no  indicator  lamps  lit  on  the 

O2NE+™/Safe-Ox+™, check that  power is 

connected and that the fuses are OK. 

2)  If the 'Good/OK' indicator is off, and the alarm indications are believed to be incorrect, 

carry  out  a  full  calibration.    If  this  fails  to  correct  the  problem  contact  your  qualified 
service engineer. 

 
A summary of the indicator lamps and buzzer operations is shown in the following tables. 

6.6.1 

Ax 

O2NE+™ 

OK LAMP 
(GREEN) 

ALARM1 LAMP 
(RED) 

ALARM2 LAMP 
(RED) 

FAULT LAMP 
(YELLOW) 

MEANING 
 

OFF 

OFF 

OFF 

OFF 

Power Off 

ON/ BLIP OFF 

OFF 

OFF 

OFF 

Normal Operation 

OFF 

FLASHING AND 
SLOW BUZZER 

OFF 

OFF 

O

2

 Level is 

< 19.5% 

 

OFF 

FLASHING 

FLASHING AND 
MED. BUZZER 

OFF 

O

2

 Level is 

< 18% 

 

OFF 

OFF 

OFF 

FLASHING AND 
SLOW BUZZER 

Calibration Error at 
Switch On 

 

OFF 

FLASHING 

OFF 

FLASHING AND 
FAST BUZZER 

O

2

 Cell Fault 

Output too High 

OFF 

FLASHING 

 

FLASHING 

 

FLASHING AND 
FAST BUZZER 

O

2

 Cell Fault 

Output too Low 

OFF 

OFF 

OFF 

FLASHING AND 
VERY FAST BUZZER 

Internal system fault 
contact Analox 

 

 

Note that alarm levels may be set at different values, depending on customer requirement 

 A calibration error or a cell fault requires the attention of a service engineer. A recalibration 

procedure may overcome the problem. 
 

♦ 

Only when Alarm 1 and Alarm 2 are enabled.

 

 

6.6.2 

Ax Safe-

Ox+™ 

OK LAMP 
(GREEN) 

ALARM1 LAMP 
(RED) 

ALARM2 LAMP 
(RED) 

FAULT LAMP 
(YELLOW) 

MEANING 
 

OFF 

OFF 

OFF 

OFF 

Power Off 

ON/ BLIP OFF 

OFF 

OFF 

OFF 

Normal Operation 

OFF 

FLASHING AND 
SLOW BUZZER 

OFF 

OFF 

O

2

 Level is 

> 23.0% 

 

OFF 

FLASHING 

FLASHING AND 
MED. BUZZER 

OFF 

O

2

 Level is 

< 19.5% 

 

OFF 

OFF 

OFF 

FLASHING AND 
SLOW BUZZER 

Calibration Error at 
Switch On 

 

OFF 

FLASHING 

OFF 

FLASHING AND 
FAST BUZZER 

O

2

 Cell Fault 

Output too High 

OFF 

FLASHING 

 

FLASHING 

 

FLASHING AND 
FAST BUZZER 

O

2

 Cell Fault 

Output too Low 

OFF 

OFF 

OFF 

FLASHING AND 
VERY FAST BUZZER 

Internal system fault 
contact Analox 

 

 

Note that alarm levels may be set at different values, depending on customer requirement 

 

 A calibration error or a cell fault requires the attention of a service engineer. A recalibration 

procedure may overcome the problem. 
 

♦ 

Only when Alarm 1 and Alarm 2 are enabled. 

Содержание O2NE+

Страница 1: ...usiness Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net O2NE Safe Ox User Manual THIS IS A PRECISION DEVICE KEEP UP RIGHT ALLOW 2 HOUR WARM UP...

Страница 2: ......

Страница 3: ...splay 6 4 2 Wall mounting 6 4 3 Wiring installation 7 4 4 AC supply models 7 4 5 DC supply models 7 4 6 Alarm repeaters 8 4 7 Alarm relay output models 10 4 8 Relay wiring 10 4 9 4 to 20mA output mode...

Страница 4: ...early to 50 relative humidity at 40 C 5 Mains voltage supply fluctuations not to exceed 10 of the nominal voltage 6 Impulse withstand over voltage category II of IEC 60364 4 443 7 Pollution degree 2 8...

Страница 5: ...nally connected equipment introducing hazardous voltages into the A50 O2NE Safe Ox or CO Clear It is the responsibility of users of this product to ensure that their equipment accessory is safe before...

Страница 6: ...On receipt of the Ax O2NE Safe Ox please check you have the following 1 Ax O2NE Safe Ox main unit and alarm repeater 2 User manual 3 Test certificate 4 Rawl plugs and screws for wall mounting 5 Drill...

Страница 7: ...s this can affect the performance of the pressure compensation sensor over time Contact Analox for installation advice if you intend to use the unit in this way 3 2 Setup Different versions of the ins...

Страница 8: ...be removed from the instrument display window This is an easy peel label and should be removed easily without leaving behind any residue 4 2 Wall mounting The Ax O2NE Safe Ox should be mounted onto a...

Страница 9: ...Safe Ox s are pre wired with a mains cable fitted with a plug suited to the destination country Where internal plug fuses are fitted these are 3 amp Ensure that the unit is connected to the correct su...

Страница 10: ...Ax O2NE Safe Ox pass through one of the cable glands on the Remote Alarm Repeater 3 Terminate the wires in accordance with the table below 1 Replace the lid of the Remote Alarm Repeater insert and ti...

Страница 11: ...ote Alarm Repeater 3 Terminate the wires in accordance with the table below 4 Connect the free wire Brown or Orange to A50 231 main PCB at JP13 terminal 5 3 Replace the lid of the Remote Alarm Repeate...

Страница 12: ...should be connected 4 8 Relay wiring The cable gland is for cables of outside diameter between 5 and 7mm if cable fitted is outside that range a suitably specified cable gland must be used Ensure tha...

Страница 13: ...Document ref P0120 820 14 August 2017 Page 11 of 23 Commercial in Confidence The customer system should be connected to this cable as follows and as per the following drawing 1 Blue wire 4 to 20mA ne...

Страница 14: ...of alarms They may be factory configured in the opposite sense if required 5 2 Alarm indications 5 2 1 Ax Safe Ox If the Ax Safe Ox detects an O2 concentration which is greater than the high alarm lev...

Страница 15: ...wn the instrument and remove from its installed location to allow access to the lid screws Remove the lid of the instrument to gain access to the internals of the instrument Disconnect the sensor from...

Страница 16: ...ses used on the Ax O2NE Safe Ox must be manufactured to a 2 tolerance Note For best performance the monitor should be powered and allowed to warm up for approximately 2 hours before calibration is per...

Страница 17: ...will illuminate sound 4 times before returning to normal operation If during the one minute wait period the red or amber LED s are illuminated consult the table below ALARM1 LAMP RED ALARM2 LAMP RED F...

Страница 18: ...w for the flow of gas over the sensor 5 After a few moments the Alarm 1 alarm should operate 6 Remove the test gas It will take a little while for the gas to diffuse out of the instrument sensor allow...

Страница 19: ...ND FAST BUZZER O2 Cell Fault Output too High OFF FLASHING FLASHING FLASHING AND FAST BUZZER O2 Cell Fault Output too Low OFF OFF OFF FLASHING AND VERY FAST BUZZER Internal system fault contact Analox...

Страница 20: ...O2NE 600g Alarm Repeater 150g Dimensions O2NE Safe Ox 175x105x75 mm Alarm Repeater 155x72x45 mm IP rating O2NE Safe Ox IP65 Alarm Repeater Hard wired IP65 Quick connect IP43 Sensor type Electro Chemi...

Страница 21: ...14 August 2017 Page 19 of 23 Commercial in Confidence 8 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Страница 22: ...O2NE Safe Ox User Manual Int Approved Document ref P0120 820 14 August 2017 Page 20 of 23 Commercial in Confidence 9 Declaration of conformity...

Страница 23: ......

Отзывы: