O2NE+ & Safe-Ox+
User Manual
Int. Approved
Document ref: P0120-820-14
August 2017
Page 9 of 23
Commercial in Confidence
4.6.3
To connect a “Strobe” Remote Alarm Repeater
1) Disconnect the power supply from the Ax
O2NE+™/Safe-Ox+™.
2) Open the Remote Alarm Repeater unit by removing the 4 screws in the front of
the case and carefully pulling the case apart. The connecting wires from the Ax
O2NE+™/Safe-Ox+™ pass through one of the cable glands on the Remote
Alarm Repeater.
3) Terminate the wires in accordance with the table below
4) Connect the free wire (Brown or Orange) to A50-231 main PCB at JP13 terminal
5.
3) Replace the lid of the Remote Alarm Repeater, insert and tighten the 4 screws, and
mount it in the desired position using the 2 fixing lugs.
4) Restore power to the Ax
O2NE+™/Safe-Ox+™. Press the switch on the Remote Alarm
Repeater once, and ensure that the four indicators and the LED strobe flash. Note that
in the presence of a genuine alarm, this test feature is disabled.
REPEATER CABLE COLOURS
Type 1
Repeater
Term
Type 2 (Cat 5)
ST1
Terminal
No
8
3
7
*See Note
4
6
4
5
5
4
6
3
7
2
8
1
1