ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 17 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
3]
Rotate the cell PCB connector clockwise to release from the cell holder.
4]
Disconnect the cell from the PCB connector.
5]
Connect the new cell to the PCB connector (Note that it will only connect in one
orientation).
6]
Fit the cell in to the housing and rotate anti-clockwise to lock in place.
7]
Refit the cell holder into the sensor bulkhead.
8]
Refit the flow adaptor into the sensor inlet.