ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 24 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
6]
Using a circular action, clean the lamp by applying light pressure to the lamp window
(do not touch the lamp window with fingers), continue polishing until a light squeaking
can be heard.
7]
Remove any residual cleaning compound from the lamp using the clean end of the
cotton bud.
8]
Ensure that the lamp is completely free from any signs of contamination prior to
refitting.
9]
Carefully place the lamp into the rear recess of the electrode stack, this will be held by
the O-ring seal in the electrode pellet.
10]
Take the electrode stack and the mated lamp and insert back into the sensor housing,
push home until the electrode stack snaps into place, the top of the electrode stack
should be flush with the top of the sensor housing.
NOTE:
FOLLOWING A VOC PID LAMP REPLACEMENT OR
CLEANING, OR REPLACING THE ELECTRODE STACK A
FULL CALIBRATION MUST BE PERFORMED. SEE P0149-
801 ACG+ USER MANUAL FOR DETAILS.