ACG+
Service & Troubleshooting Manual
Int. Approved
Document ref: P0149-825-00
October 2022
Page 13 of 37
Copyright © 2022 Analox Ltd. All Rights Reserved.
Commercial in Confidence
4.1.1
Oxygen cell replacement
The part number for your sensor’s replacement cell is:
9100-9212-9HM
When a replacement cell is received check that the cell has not leaked. The cells are themselves sealed
and do not under normal circumstances present a health hazard however if leakage of the electrolyte
has occurred use rubber gloves and wear chemical splash goggles to handle and clean up. Rinse
contaminated surfaces with water. If contact is made with the electrolyte, refer to section 1 Safety
Information.
1]
Remove the flow adaptor from the sensor inlet.
2]
Loosen the corner screws of the enclosure and remove the lid.