To remove all visible soil, wipe down the pump with non-abrasive
paper towel, sponge, or lint free soft cloth moistened with cleaning
solution.
•
To remove soil from crevices, use a soft brush.
•
After the first cleaning, replace it with a new wet wipe and clean it
again.
3. Cleaning Frequency
•
It is recommended that the pump is cleaned to eliminate stains and
soils after each use or at least every day.
•
A cleaning is also necessary when the pump is allocated to a new
patient.
4. Pump Housing Cleaning
•
Refer to General Cleaning Directions before starting.
•
Clean the outside surface with a solution of warm water and a mild
dishwashing detergent and a non-abrasive sponge, paper towel, or
soft cloth.
5. Power Adapter Cleaning
•
Refer to General Cleaning Directions before starting.
•
The power adapter normally does not require cleaning. If soiling of the
power adapter is observed, unplug from outlet and wipe the exterior
surfaces of the power adapter with a non-abrasive dry or slightly damp
cloth.
•
Allow excess moisture to evaporate from the power adapter prior to
use.
6. Rotor Cleaning
•
Refer to General Cleaning Directions before starting.
•
Open the door.
•
Use a cotton swab to clean the rollers thoroughly with a solution of
warm water and a mild dishwashing detergent.
7. Sensors Cleaning
•
Open the door.
•
Use a damp cotton swab to clean the sensors and the pathway of the
feeding set.
12.3 Disinfection
1. Disinfection Chemicals
The pump can be disinfected with the following solution:
•
Alcohols, ammonium salt disinfectants can be used for disinfection of the
product.
•
To wet the devices, use wetted lint free wipes and wipe as necessary to
maintain visual wetness for 5 minutes.
2. Disinfection Frequency