File
Name
IFU of
AMSure
®
Enteral
Feeding
Pump and Feeding Set
REF
#
L10196_US
Version
B
Page
18
of
81
AMSINO Confidential
3
Single Dose
Indicate number of single feeding
dose
4
Feed Count
– In Setting
Status
Indicate number of feedings
5
Single Fed
Indicate volume delivered in current
single feeding task
6
Interval time
– In
Setting Status
Interval time to be set between
feedings
Note: the maximum interval time is
23 h 59 mins.
Single Time Left
– In
Running/Stop Status
Indicate time left to feeding end of
current single feeding task
7
-
Total Fed
Indicate total fed volume delivered
to the patient
8
-
Feed Count
– In
Running/Stop Status
Indicate number of feeds that have
been run from total programed
feeds.
9
KTO Running
The icon is rotating to show KTO
running status. Refer to 7.2.8 KTO
10
Prime
Enter prime page. Refer to 7.2.2 for
Prime section.
11
New Feeding
End current feeding task and goes
to setting status.
12
Lock
Execute screen lock function. Refer
to 7.2.9 Screen lock
13
Unlock
Execute screen unlock function.
Refer to 7.2.9 Screen lock
14
KTO
Execute KTO function. Refer to
7.2.8 KTO
15
Back
Back to previous page.
Note: In all programmed feeds in one intermittent setting, all feeding
parameters including feeding rate, single feeding dose, interval waiting time
are same.
7. Operation Guidance
Caution
All the enteral feeding sets are disposable. The operation steps are as
following: install the pump → switch on with self-testing → install the feeding
set→ RFID tag swipe→ set parameters → priming→ connect to patient’s