Enter "System Maintenance" page from the home page, which can provide the
functions such as calibration, RFID switch on/off, sensor parameter viewing,
system self-test, and factory setting restoration. Maintenance engineers should
provide the password before entering "System Maintenance".
Fig 7.3.5 System Maintenance
7.3.5.1 Calibration
This pump can only be used with dedicated enteral feeding set. The pump has
been calibrated before leaving the factory and can be used directly. To keep
normal use, re-calibrate the pump every 12 months.
Carry out accuracy calibration according to the following steps:
1. Items Required for Calibration
The calibration shall be carried out at room temperature.
No.
Items
Model
Quantity(pcs)
Remark
1
Enteral feeding pump
ASEP00
1
2
Enteral feeding set
ENF-1004K2
1
3
Feeding container
/
4
Water
Distilled water
1
Density (g/mL)
0.998
5
I.V. pole
/
1
6
Electronic balance
/
1
Accuracy:0.01g
7
Collecting container
About 10mL
1
Note:
1) No model specified for feeding container, IV pole and electronic
balance.
2) ENF-1004K2 is selected as representative. User also can choose
other enteral feeding set in the list (refer to 9. Dedicated Enteral
Feeding Sets (Disposable))
2. Preparation for Calibration