![Amsino AMSure Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/amsino/amsure/amsure_user-manual_2936262043.webp)
File
Name
IFU of
AMSure
®
Enteral
Feeding
Pump and Feeding Set
REF
#
L10196_US
Version
B
Page
43
of
81
AMSINO Confidential
Fig 7.3.1 Feed History
A maximum of 72 hours of feeding
history can be retrieved, excluding
data from the current hour.
Input hour number (1-72) in the
upper "History" window by touching
the
orange
number to check the fed
volume during the time range.
The Total Fed volume from the time
when the data was cleared. The data
is real-time record.
It’s normally cleared for a new
patient or before a new feeding.
Note: The 72-hour history includes time when the pump is powered off.
7.3.2 Log Record
Touch
“Log Record” button from the home page to query operation and alarm
records. The pump can save approximately 10,240 logs. When the memory is
full, the latest 2048 logs still will be saved. but earlier log records will be
deleted so that the pump can keeps saving the latest log records.