AS5000 Programmer OPERATIONAL MANUAL
Revision 1.10, 04-March-2010
Page 12 of 24
www.austriamicrosystems.com
-
Zero Position field: Set button writes the actual angle value into the Zero Position register of the encoder. This
programming is not permanent. The actual value will be 0 after zero position programming. To reset the zero
position register, or to set a new zero position, click on Reset first.
-
Read interval is the SSI stream readout and refresh rate the GUI.
Figure 14: OTP tab of AS5045
•
AS5245:
Load
reads the OTP configuration from the permanent fuse register.
Read
reads the OTP configuration from the volatile register.
Write
writes the bits checked by the user into the volatile OTP register. Selecting the SSI after the configuration is
possible to check the effect.
Zap!
writes permanently the checked OTP bits to the encoder’s fuses.
Selected die
(dual die AS52xx devices only)