TECHNICAL SPECIFICATIONS
44
D-0115696-C (OM038)
– Otowave 302 Operating Manual
Printing
Supported printer:
Sanibel MPT-II
Interface:
Cable supplied
Information printed:
Tympanogram, Tympanogram analysis parameters,
Reflex graphs, Reflex analysis parameters, Serial
Number of device, Last and Next Due Calibration
dates; space for patient & clinician’s details to be
entered.
PC Interface
Serial Interface:
USB Version 1.1
Information sent:
Patient header, full left & right ear data.
7.3.
TYMPANOMETRY
Probe Tone
Frequency:
226 Hz ±2% and 1000 Hz ±2%
Level:
85 dB SPL ±2dB and 79 dB SPL ±2 dB over ECV
range
Pressure
Range:
+200 daPa to -400 daPa ±10daPa or ±10 %
(whichever is larger) over range 0.1 ml to 6 ml
Limits (safety cutout):
+600 and -800 daPa
Sweep
Speed:
Selectable: 100, 200 or 300 daPa/sec
Analysis
Admittance peak level (in ml or
m
Ʊ
) pressure at peak;
Gradient in daPa (for 226Hz and 1000Hz);
Ear Canal Volume (ECV) @ 200 daPa or -400 daPa
Number of samples
stored
100 per tympanogram
7.4.
ACOUSTIC REFLEX TESTING
Ipsilateral
Test Frequencies:
500 Hz, 1 kHz, 2 kHz & 4 kHz (±2 %)
Level:
Configurable over range: 500Hz, 1kHz, 2kHz & 4kHz
(+/-2%)
70dBHL to100dBHL (+/-3dB)
(2kHz level is restricted to maximum 95dBHL for ear
canal volumes greater than ~3.5ml)
(4kHz level is restricted to maximum 85dBHL for ear
canal volumes greater than ~3.5ml & maximum 95dBHL
for all ear canal volumes)
Number of reflex levels
presented below the
selected maximum and
step size(s) available:
100dBHL max, with 5dB or 10dB steps
95/90/85dBHL max, with 5dB steps
Contralateral
Test Frequencies:
500Hz, 1kHz, 2kHz & 4kHz (±2%)
Level:
Configurable over range:
500Hz, 1kHz, 2kHz & 4kHz
(+/-2%)
70dBHL to110dBHL (+/-3dB)
(1kHz level is restricted to minimum 75dBHL for ear
canal volumes less than ~0.2ml)